Safety and quick start guide for the Druck DPI612 portable pressure calibrator, covering setup, power modes, dashboard navigation, and specifications.
The DPI612 is a battery-powered instrument for electrical measure and source operations. It measures and displays, via user-interchangeable pressure modules, pneumatic or hydraulic pressure/vacuum generated by the instrument, with the type and range of pressure generation dependent on the selected model (DPI612-PFX, DPI612-PFP, or DPI612-HFP).
It can be powered from a rechargeable LiIon battery pack or 8 AA alkaline batteries, and its dashboard provides quick selection of functions such as the calibrator application. The guide also covers safety warnings, parts identification, external pressure connections, power modes, and general specifications.
| Operating temperature range | -10 °C to +50 °C (+14 °F to +122 °F) |
|---|---|
| Optimized calibration temperature | +10 °C to +30 °C |
| Operating temperature with optional mains power supply | 0 °C to 40 °C |
| Storage temperature | -20 °C to +70 °C (-4 °F to +158 °F) |
| Humidity | 0 to 90 % RH, non-condensing |
| Power supply | 100 – 260 V ±10 %, 50/60 Hz AC |
| Altitude | Up to 2000 m |
| Rechargeable battery | LiIon 14.1 Wh, GE part CC3800GE, 3.7 V, 3800 mAh |
| Rechargeable battery charge temperature | 0 °C to 40 °C (32 °F to 104 °F) |
| Rechargeable battery discharge temperature | -10 °C to 60 °C (14 °F to 140 °F) |
| Non-rechargeable batteries | 8 × AA alkaline |
| Pressure leak rate, DPI612-PFX | 0.01 bar/min at 20 bar |
| Pressure leak rate, DPI612-PFP | 0.02 bar/min at 100 bar |
| Pressure leak rate, DPI612-HFP | 1 bar/min at 1000 bar |
| Vacuum leak rate, DPI612-PFX/PFP | 0.005 bar/min at 95 % vacuum |
| Recommended pressure module, DPI612-PFX | 2 to 20 bar (3 to 300 psi) |
| Ingress protection | IP54 |
| Display | LCD color display with touch screen |
| Certification | CE marking |
Momentarily press the power button until the GE logo appears.
Press and release the power button, then select SWITCH OFF from the POWERDOWN OPTIONS window; or press and hold the power button until the screen is blank.
It uses a rechargeable LiIon battery pack (GE part CC3800GE) or 8 AA alkaline batteries.
2 to 20 bar (3 to 300 psi).