User manual for DESCO's 19787 Digital Surface Resistance Meter Kit, covering installation, operation, maintenance, and calibration for ESD resistance…
This manual provides detailed instructions for installing, operating, and maintaining the DESCO 19787 Digital Surface Resistance Meter Kit. It explains how to measure surface resistance point-to-point (Rtt) or to ground (Rtg) in accordance with ESD Association standards. The kit includes features like temperature and humidity measurement, adjustable test voltages (10V or 100V), and exponent LEDs for quick resistance reading. It also covers electrode placement, test procedures, and reporting requirements for compliance with ANSI/ESD standards.
| Resistance Measurement Accuracy | ±10% (±20% for values ≥1 x 10¹¹ ohms) |
|---|---|
| Resistance Range | <1 x 10³ ohms to >1 x 10¹² ohms |
| Under Load Voltages | 10 and 100 volts ±5% |
| Electrification Period | Approximately 15 seconds |
Clip the black test lead to a grounded point, place one 5-pound electrode on the surface, and push and hold the Test Button until the mantissa displays.
Use 100 volts for materials with resistance ≥1 x 10⁶ ohms.
The display shows mantissa and exponent values; for example, #8 exponent LED + "7.14" = 7.14 x 10⁸ ohms.