DESCO 19787 Digital Surface Resistance Meter Kit User Manual
This manual provides detailed instructions for installing, operating, and maintaining the DESCO 19787 Digital Surface Resistance Meter Kit. It explains how to measure surface resistance point-to-point (Rtt) or to ground (Rtg) in accordance with ESD Association standards. The kit includes features like temperature and humidity measurement, adjustable test voltages (10V or 100V), and exponent LEDs for quick resistance reading. It also covers electrode placement, test procedures, and reporting requirements for compliance with ANSI/ESD standards.
Read onlineDESCO 19787 specifications
| Resistance Measurement Accuracy | ±10% (±20% for values ≥1 x 10¹¹ ohms) |
|---|---|
| Resistance Range | <1 x 10³ ohms to >1 x 10¹² ohms |
| Under Load Voltages | 10 and 100 volts ±5% |
| Electrification Period | Approximately 15 seconds |
Source: DESCO 19787 Digital Surface Resistance Meter Kit User Manual
All DESCO 19787 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
How do I measure surface resistance to ground (Rtg)?
Clip the black test lead to a grounded point, place one 5-pound electrode on the surface, and push and hold the Test Button until the mantissa displays.
From: DESCO 19787 Digital Surface Resistance Meter Kit User Manual
What test voltage should I use for materials with resistance ≥1 x 10⁶ ohms?
Use 100 volts for materials with resistance ≥1 x 10⁶ ohms.
From: DESCO 19787 Digital Surface Resistance Meter Kit User Manual
How should hammer electrodes be placed for floor testing?
For floor testing, the two hammer electrodes must be placed at a distance of at least 91.5 cm apart.
How do I convert the shielding bag volume resistance result?
To convert the result to volume resistivity, multiply the test result by 10.