This manual provides instructions for the Bionime GM260 blood glucose meter, covering setup, testing procedures, quality control, and troubleshooting.
The Bionime GM260 blood glucose meter is designed for both self-testing and professional use. It automatically detects coding and measures blood glucose levels using fresh whole blood obtained via a finger prick or other capillary sites. The device is compatible only with the GS260 test strips and specific control solutions to ensure accurate results.
This guide details the product components, including the meter, test strips, and carrying case. It explains how to start the device, replace the CR2032 battery, and configure settings such as date, time, and volume. The manual also covers the testing procedure, including non-fingerstick testing (AST), interpreting results, and performing quality control tests. Additionally, it lists fault messages and provides troubleshooting steps for common issues like sampling errors or test strip errors. This manual is written in Chinese.
| Blood Volume Required | 1.0 µL |
|---|---|
| Measurement Range | 0.6-33.3mmol/L (10-600 mg/dL) |
| Operating Temperature | 10-40℃ (50-104 ℉) |
| Storage Temperature | -10~60℃ (14~140 ℉) |
| Battery Type | 1 CR2032 |
| Battery Life | Approx. 1000 tests |
| Memory Capacity | 500 results |
| Dimensions | 98mm x 45mm x 16.5mm |
| Weight | 55.0 ± 5g (with battery) |
The minimum blood sample volume required is 1.0 µL.
The meter can store up to 500 test results, including date and time.
A new battery can provide power for approximately 1000 tests.
No, the meter is not approved for use on newborn infants.