Installation and operating instructions for Ashcroft AVC-1000 and AVC-3000 pressure volume controllers, including a parts list.
This document gives installation and operating instructions for the Ashcroft Pressure Volume Controllers, models AVC-1000 and AVC-3000. Installation covers fitting a valve to the vent port and to the inlet port on the pressure balance valve assembly, fitting a tee to the rear of the AVC unit, and connecting a pressure standard and the device to be calibrated or tested to the tee. All plumbing connections are made to 1/8NPTF ports.
The operation section walks through opening the pressure balance valve and the vent valve, setting the control knob, closing the vent valve, bringing the system to the desired pressure through the inlet valve, and then increasing or decreasing system pressure with the control knob. A note states that the pressure balance valve opens automatically if the pressure across the piston exceeds safe levels, factory set at a nominal pressure differential of 150 psi. A parts list with item numbers, part numbers and descriptions for the 1000 psi and 3000 psi units follows.
| Plumbing connections | 1/8NPTF ports |
|---|---|
| Pressure balance valve relief setting | Factory set at a nominal pressure differential of 150 psi |
| Model variants | 1000 psi unit and 3000 psi unit |
| Optional table mount bracket part number | 621X001-02 |
| Panel mounting kit part number | 617X006-01 |
It opens automatically if the pressure across the piston exceeds safe levels; this is factory set at a nominal pressure differential of 150 psi.
All plumbing connections are made to 1/8NPTF ports.
Turn the control knob clockwise to increase system pressure or counterclockwise to decrease system pressure.
The parts list gives separate part numbers for the 1000 psi and 3000 psi units, such as the actuator, lead screw, O-ring, piston and label, and notes that some quantities are 2 in model AVC-3000.