Armstrong Steam QM-3 Steam Quality Monitor Installation and Operation Manual

Chinese installation and operation manual (IOM-245-V2.0-CN) for the Armstrong Steam QM-3 steam quality monitor, covering setup, software, and troubleshooting.


Armstrong Steam QM-3 Steam Quality Monitor Installation and Operation Manual - cover page
Brand
Armstrong
Category
Monitoring Equipment
Document type
User Manual
Language
Chinese
Pages
32
File format
PDF
File size
4.2 MB
Published
10 November, 2026
Updated
10 November, 2026
MD5 checksum
19B711CFF7E5919ED0ECFCCE35925BDE

About this manual

This is the Chinese installation and operation manual (document IOM-245-V2.0-CN) for the Armstrong Steam QM-3 steam quality monitor. The unit is designed to replace manual pure steam testing and provide real-time data demonstrating that steam quality meets regulatory requirements. It is set up and calibrated to test the parameters defined in the EN285 standard, performing three tests: steam dryness fraction, superheat, and non-condensable gas (NCG) content.

The manual describes the calorimeter assembly and cabinet, gives technical specifications, and walks through site selection, typical installation, and startup steps. It also covers software navigation (standard and special menus), troubleshooting, software updates, a parts list, wiring diagrams, schematic drawings, Modbus communication connections, and the limited warranty. Measured data can be recorded to a data logger and transmitted via Modbus. This manual is written in Chinese.

What's inside

  • 修订记录
  • 安全性
  • 缩略词表
  • 产品描述
  • 规格
  • 安装
  • 启动步骤
  • 软件导航
  • 故障排查
  • 组件和零部件清单

Specifications

Calorimeter operating temperature range (T1)0–150 °C (32–302 °F)
Calorimeter maximum allowable temperature (T1)180 °C (356 °F)
Saturated steam operating pressure range (P)0.5–4 barg (7–60 psig)
Superheated steam operating temperature range (T2)0–180 °C (32–356 °F)
Steam dryness fraction0.85–1, display accuracy ± 0.01
Non-condensable gas content≤ 15%
Condensate temperature range (T3)0–80 °C (32–176 °F), operating 65 °C (149 °F)
Expected steam consumption1.5 kg/h (3.3 lbs/h) @ 3 barg (45 psig)
Expected water consumption15 L/h (4 gal/h) @ 10 °C (50 °F)
Electrical requirements115/230 VAC, 50/60 Hz, 100 W
WeightsCalorimeter assembly approx. 5 kg (11 lbs); cabinet approx. 19 kg (42 lbs)

Frequently asked questions

What tests does the Steam QM-3 perform?

It performs three tests: calculation of steam dryness, calculation of superheat, and measurement of non-condensable gas (NCG) content. The NCG measurement runs first, and if the result is within the normal range, the dryness and superheat measurements are carried out.

Which standard is the Steam QM-3 calibrated for?

The unit is set up and calibrated to test the parameters specified in the EN285 standard.

Can measurement data be exported from the Steam QM-3?

Yes, data measured by the Steam QM-3 can be recorded to a data logger and transmitted via Modbus; connection information is given in Appendix 3 of the manual.

Can the calorimeter and cabinet be swapped between units?

No, the calorimeter assembly and the cabinet are matched equipment and must not be interchanged. If a cabinet has been connected to a different calorimeter, it must be calibrated before use.