User's guide for the Anritsu Site Master S810D/S820D handheld cable and antenna analyzers: setup, measurements, power monitoring and reference data.
This user's guide covers the Anritsu Site Master models S810D and S820D, handheld cable and antenna analyzers for measuring SWR, return loss, cable insertion loss and for locating faulty RF components in antenna systems. The S810D covers 0.025 to 10.5 GHz and the S820D covers 0.025 to 20 GHz. The instrument has a keypad for data entry, a TFT LCD for graphic indications of measurements, and runs for up to 1.5 hours on a fully charged field-replaceable battery or from a 12.5 Vdc source.
The guide opens with warranty, trademark and notice statements and a table of contents, then covers options and accessories, performance specifications, preventive maintenance and calibration, front panel and soft key functions, self test and error messages, battery handling, getting started procedures, cable and antenna analyzer measurements, power measurement, and the Anritsu Tool Box with Line Sweep Tools software. Appendices provide coaxial cable and waveguide reference data, windowing, and signal standards.
| Frequency Range (S810D) | 0.025 to 10.5 GHz |
|---|---|
| Frequency Range (S820D) | 0.025 to 20.0 GHz |
| Frequency Resolution | 10 kHz (100 kHz for Distance To Fault) |
| Output Power (RF Out port) | < 0 dBm at any particular frequency |
| Number of Data Points | 130, 259, 517 |
| Return Loss Range | 0.00 to 60.00 dB (resolution 0.01 dB) |
| VSWR Range | 1.00 to 65.53 (resolution 0.01) |
| Coax/Waveguide Insertion Loss Range | 0.00 to 30.00 dB (resolution 0.01 dB) |
| Battery Operating Time | Up to 1.5 hours of continuous operation from a fully charged field-replaceable battery |
| Measurement Speed | ±2 sec/sweep for 517 data points (CW on) |
The S810D covers 0.025 to 10.5 GHz, while the S820D covers 0.025 to 20.0 GHz.
It is capable of up to 1.5 hours of continuous operation from a fully charged field-replaceable battery.
Option 22NF provides 2-port cable-loss mode and includes an external N(f) CW source module CWM220B-NF, a 560-7N50B detector and a 66379 adaptor cable.
A warm-up time of 15 minutes should be allowed before verifying system specifications.