Quick operation instructions for the Alicat 21V1 handheld gas mass flow meter, covering menu settings, startup checks, serial commands and troubleshooting.
The document is a two-page Chinese quick operation guide for the Alicat 21V1 handheld gas mass flow meter. It explains menu settings for flow and pressure averaging, serial communication, standard conditions, gas selection and engineering units.
It also gives startup checks, process connection and gas filtration guidance, key functions, power requirements, and troubleshooting for key lock, inaccurate flow, computer communication and display brightness. This manual is written in Chinese.
| Flow range | 0.5 SCCM — 6000 SLPM |
|---|---|
| Measurement range | 0.01~100% of full scale |
| Response time | better than 10 ms |
| Built-in gases | 98 gases |
| Power supply | 5 VDC (Micro-USB-B) |
| Charging time | 4 hours when powered off |
| Battery life | 18 hours monochrome, 5 hours color screen |
Use MENU-SETUP-Active Gas-Standard, move the arrow to the desired gas and press SET.
LPM and SLPM should read 0; if they do not, perform zeroing.
Wrap 3–5 layers of PTFE tape starting from the third thread.
It means the keys are locked; clear it by entering the unlock command through digital communication.