This document is a simple guide for operation and troubleshooting of the ALICAT 21V1 gas mass flow meter. It provides instructions on how to modify flow and pressure averaging, as well as common serial commands. It also offers troubleshooting methods for issues such as non-working buttons, inaccurate flow measurement, and communication problems with a computer.
This document serves as a quick operation guide and troubleshooting reference for the ALICAT 21V1 Gas Mass Flow Meter. It covers essential procedures including pre-use checks, installation, operation, calibration, and serial communication parameter settings. The guide also details how to modify flow averaging, pressure averaging, and standard conditions via the device menu. This manual is written in Chinese.
| Flow Range | 0.01~100% Full Scale |
|---|---|
| Response Time | Better than 10 ms |
| Power Supply | 7-30 VDC |
| Max Static Pressure (20/20S/62) | 160 psig |
| Max Differential Pressure (20/20S/62) | 75 psid |
| Max Static Pressure (20W/62W) | 65 psig |
| Max Differential Pressure (20W/62W) | 10 psid |
| Max Static Pressure (20Q/62Q) | 385 psig |
| Max Differential Pressure (20Q/62Q) | 75 psid |
Go to MENU- SETUP-Sensor and select Flow Averaging to adjust the value between 1-255 ms for smoother display during fluctuations.
The device may be locked; you need to unlock it via digital communication by inputting the unlock command.
Check if the displayed pressure and temperature are normal; if not, the sensor likely needs repair.
It is recommended to recalibrate the flow meter once a year, or return it for calibration if more than a year has passed.