The TPS 334 is a thermopile sensor housed in a TO‑5 type package with a 0.7 × 0.7 mm² absorber area and a 30 kΩ thermistor temperature reference. It is designed for temperature measurement applications such as pyrometers and ear thermometers and can be equipped with a G9 (8‑14 mm) filter for precision remote sensing.

The datasheet details absolute maximum ratings, electrical parameters (responsivity 55 V/W, time constant 25 ms, noise voltage 35 nV/√Hz, NEP 0.64 nW/√Hz, detectivity 1.1 × 10⁸ cm√Hz/W), mechanical drawing, filter characteristics, and handling precautions including ESD protection and soldering limits. Quality is ensured by ISO‑9001 certification and lot acceptance testing.

What's inside

  • Scope
  • General Characteristics
  • Electrical Parameters
  • Optical Characteristics
  • Mechanical Drawing
  • Quality System
  • Lot Acceptance Test
  • Handling Requirements

Specifications

Ambient Temperature Range-40 °C to 100 °C
Absorber Resistance (R_TP)50 kΩ – 100 kΩ (typical 75 kΩ)
Responsivity (S)55 V/W
Time Constant (τ)25 ms
Noise Voltage (V_RMS)35 nV/√Hz
Noise Equivalent Power (NEP)0.64 nW/√Hz
Detectivity (D*)1.1 × 10⁸ cm√Hz/W
Field of View7 ° – 10 °

Frequently asked questions

What is the ambient temperature range for the TPS 334?

The TPS 334 operates from –40 °C up to 100 °C.

How should the TPS 334 be handled to avoid damage?

Avoid electrostatic discharge, do not expose the sensor to aggressive detergents, clean the window with alcohol and a cotton swab, and limit hand or wave soldering to a maximum of 300 °C for less than 10 seconds.

What filter can be used for precision remote temperature sensing with the TPS 334?

A G9 (8 mm to 14 mm) long‑pass infrared filter can be equipped for precision remote temperature sensing.

What is the responsivity of the TPS 334 thermopile sensor?

The sensor has a responsivity of 55 V/W (typical, without filter or optics, at 500 K, 1 Hz).