The NIOX VERO® Respiratory Inflammation Monitor User Manual provides information on the device's specifications, instructions for use, precautions, and more.
The document is the Spanish user manual for the NIOX VERO respiratory inflammation monitor, published by Circassia AB in September 2019. It covers essential information for healthcare professionals, including device components, installation, user interface navigation, and step-by-step procedures for measuring fractional exhaled nitric oxide (FeNO) and environmental NO.
The manual details specific operational constraints and safety warnings, such as the maximum number of exhalations per hour to prevent internal condensation and the prohibition of using alcohol-based substances near the device. It also outlines the device's technical specifications, including dimensions, electrical characteristics, and battery autonomy, as well as procedures for connecting the device to a PC via USB or Bluetooth using the NIOX Panel software. This manual is written in Spanish.
| Operating Temperature | +10°C to +35°C |
|---|---|
| Operating Humidity | 20% to 80% HR |
| Maximum Continuous Exhalations | 10 per hour |
| Device Lifespan | 5 years or 15,000 measurements |
| Sensor Lifespan | 12 months |
| Software Version | 1D2x-xxxx and 1E2x-xxxx |
Up to 10 exhalations per hour continuously, or 20 exhalations in an hour if the device is paused for at least 30 minutes.
No, substances containing alcohol must not be used near or on the device as they can cause incorrect measurement results.
The device has a maximum lifespan of 5 years of use, 15,000 measurements, or until the expiration date, whichever comes first.