Non-contact infrared thermometer instruction manual. This product uses an ambient temperature of 15°C to 35°C, with an optimum temperature of 25°C. Do not place the product near a charged object to avoid electric shock. Do not use the product in an environment with a relative humidity greater than 85%. Do not place the product too close to an electromagnetic range (such as a radio or mobile phone). Do not place the product in direct sunlight or near a heat source, and do not allow it to come into contact with water. Do not hit or drop the product. If it is damaged, do not use it. Hair, sweat, hats or scarves on the forehead will affect the accuracy of the measurement data. Please ensure that the measurement distance is between 1cm and 5cm. In order to obtain accurate and stable measurement data, when the room temperature changes greatly, the measuring instrument should be placed indoors for 30 minutes before use. When cleaning is required, please use alcohol to gently wipe the surface of the instrument. If the product has any problems, please contact the distributor and do not attempt to repair it yourself.
This manual is for the UT30H non-contact infrared forehead thermometer, intended for measuring the body temperature of children over 5 years old and adults by detecting forehead thermal radiation. It covers safety rules, main structure, intended use, contraindications, operating and storage conditions, technical description, operation guidance, calibration mode conversion, maintenance, storage precautions, waste disposal, service life, accessories, warranty and electromagnetic compatibility.
The documented technical parameters include a human-mode measurement range of 32.0-45.0 C, accuracy of ±0.2 C for 35.0-42.0 C and ±0.3 C for 32.0-34.9 C and 42.1-45.0 C, 0.1 C resolution, LCD display, last-measurement memory, fever alarm above 37.2 C, automatic shutoff after about 15 seconds, two 1.5 V AAA batteries, dimensions of 140.7 x 89.7 x 35.9 mm, weight of 98.7 g, clinical repeatability not exceeding 0.3 C, clinical accuracy of 0.08 C, operating and storage environmental limits, a 3-year service life and a 1-year warranty. This manual is written in Chinese.
| Measurement range (human mode) | 32.0-45.0 C (89.6-113.0 F) |
|---|---|
| Measurement accuracy | ±0.2 C/0.4 F for 35.0-42.0 C; ±0.3 C/0.5 F for 32.0-34.9 C and 42.1-45.0 C |
| Display | LCD, 3.5-digit with special icons |
| Memory | Last measurement |
| Fever alarm | Above 37.2 C (98.9 F), buzzer long beep plus two beeps |
| Automatic shutoff | About 15 seconds without operation |
| Battery | 2 x 1.5 V AAA (zinc-manganese) |
| Dimensions | 140.7 x 89.7 x 35.9 mm |
| Clinical repeatability | Not more than 0.3 C (0.5 F) |
| Operating humidity | Up to 85% RH, no condensation |
| Operating pressure | 700-1060 hPa |
| Storage and transport humidity | Up to 93% RH, no condensation |
| Service life | 3 years |
| Warranty | 1 year from purchase |
| Measurement distance | 1-5 cm |
| Intended users | Children over 5 years old and adults |
| Ingress protection | IPXO |
Keep the measurement distance between 1 and 5 cm.
The product has a one-year free warranty from the date of purchase.
It uses two 1.5 V AAA zinc-manganese batteries.
It turns off automatically after about 15 seconds without operation.