This manual describes the features, specifications, design, operation methods, etc. of the high voltage insulation resistance tester KEW3125A/3025A.
This manual is for the Kyoritsu KEW3125A/3025A high-voltage insulation resistance tester. It describes safety precautions, applicable standards, instrument layout, LCD display, battery checks, and measurement procedures.
Features include automatic discharge of capacitive circuits, a backlight, audible and visible active-circuit warnings, automatic power off after 10 minutes, and automatic PI and DAR measurement. It also covers battery replacement, accessories, and product disposal. This manual is written in Spanish.
| Applicable standards | IEC 61010-1, -2-030; IEC 61010-031; IEC 61326-1, -2-2; IEC 60529 IP40; ES 50581 RoHS Directive |
|---|---|
| Measurement category | CAT III 600 V / CAT IV 300 V |
| Pollution degree | 2 |
| Ingress protection | IP40 |
| Nominal test voltages | 250 V, 500 V, 1000 V, 2500 V |
| Accuracy | +5% of reading +3 digits |
| Auto power off | 10 minutes |
Yes. Charges stored in capacitive circuits are discharged automatically after measurement.
It activates if there is no function change or button press for 10 minutes.
It is designed to comply with IEC 61010-1, -2-030 and IEC 61010-031.
The instrument has an IP40 rating.