Kyoritsu 3165/3166 Insulation Resistance Tester User Manual

This document is about how to use a digital multimeter.


Kyoritsu 3165/3166 Insulation Resistance Tester User Manual - cover page
Brand
KYORITSU
Category
multimeter
Document type
User Manual
Language
Spanish
Pages
3
File format
PDF
File size
480 KB
Published
17 November, 2023
Updated
17 November, 2023
MD5 checksum
C621D10F621E1A3BCC414D000DA3F1BF

About this manual

This manual covers safety precautions, features, specifications, operation, battery replacement, case and accessory handling, and service and calibration for the Kyoritsu 3165/3166 insulation resistance tester.

It includes measurement procedures for insulation resistance and AC voltage, plus model-specific ranges, accuracy, overload protection, and physical specifications. This manual is written in Spanish.

What's inside

  • Safety precautions
  • Features
  • Specifications
  • Preparation for tests
  • AC voltage warning function
  • Insulation resistance measurement
  • Battery replacement
  • Case and accessories
  • Service and calibration

Specifications

Test voltage500 V (Model 3165), 1000 V (Model 3166)
Insulation resistance range1000 MΩ (Model 3165), 2000 MΩ (Model 3166)
Mid-scale value20 MΩ (Model 3165), 50 MΩ (Model 3166)
Open-circuit output voltage+20% / -0% of test voltage
Short-circuit output current230 µA max (Model 3165), 450 µA max (Model 3166)
Accuracy, 1-500 MΩ±5% of reading
Accuracy, 2-1000 MΩ±5% of reading
Accuracy, other ranges±0.7% of scale length
AC voltage warning range0-600 V AC
AC voltage warning accuracy±3% of scale length
Power source4 AA batteries or equivalent
Overload protection, insulation resistance range, Model 3165600 V (DC + AC p-p) for 10 sec
Overload protection, insulation resistance range, Model 31661200 V (DC + AC p-p) for 10 sec
AC voltage range overload protection720 V (DC + AC p-p) for 10 sec
Operating temperature and humidity0-40°C, relative humidity 85%
Withstand voltage3700 V AC for one minute
Insulation resistance of housingMore than 50 MΩ at 1000 V
Dimensions90 x 137 x 40 mm
Weight330 g
AccessoriesTest probes Mod. 7025 (1 set), probe pouch, strap, R6 batteries (4 pcs), manual

Frequently asked questions

How many batteries does the tester use?

It uses four 1.5 V AA (R6) batteries or equivalent.

What is the operating temperature and humidity range?

It operates from 0 to 40°C with relative humidity up to 85%.

What is the overload protection for Model 3165?

The insulation resistance range is protected for 600 V (DC + AC p-p) for 10 seconds.

What are the dimensions and weight?

The tester measures 90 x 137 x 40 mm and weighs 330 g.

Download PDF (480 KB)

More KYORITSU Manuals


View all KYORITSU manuals →