GB/T 26098-2010 Roundness Measuring Instrument Standard

This document specifies the terminology and definitions, types and basic parameters, requirements, test methods, marks and packaging of roundness measuring instruments.


GB/T 26098-2010 Roundness Measuring Instrument Standard - cover page
Category
Other products
Document type
Other
Language
Chinese
Pages
8
File format
PDF
File size
329 KB
Published
16 November, 2023
Updated
16 November, 2023
MD5 checksum
34DB40BE60D0954B816C3C2ABAF8417B

About this manual

This document is the Chinese national standard GB/T 26098-2010 for roundness measuring instruments. It defines terms, instrument types, basic parameters, requirements, test methods, and marking and packaging.

It applies to all roundness measuring instruments and includes an informative appendix on the roundness multiple rotational error separation method. This manual is written in Chinese.

What's inside

  • Scope
  • Normative references
  • Terms and definitions
  • Type and basic parameters
  • Requirements
  • Test methods
  • Marking and packaging
  • Appendix A: roundness multiple rotational error separation method

Specifications

Instrument typesTurntable (spindle) rotating type and sensor (spindle) rotating type
Main parametersMaximum measurable diameter, maximum measurable height, and magnification
Transmission band ranges(1-15), (1-50), (1-150), (1-500), (1-1500), (15-500), and (15-1500) UPR
Stylus tip radius series0.25 mm, 0.8 mm, 2.5 mm, 8 mm, and 25 mm
Measuring force rangeAdjustable within (0-0.25) N
Amplifier conversion errorAdjacent bands not greater than 3%; any band relative to the calibration band not greater than 5%
Calibration errorNot greater than 1%
StabilityMeasurement system indication stability not greater than 2%

Frequently asked questions

What types of roundness measuring instruments does the standard cover?

It covers turntable (spindle) rotating type and sensor (spindle) rotating type instruments.

What transmission band ranges are specified for roundness profiles?

The specified ranges are (1-15), (1-50), (1-150), (1-500), (1-1500), (15-500), and (15-1500) UPR.

What measuring force range should a roundness measuring instrument provide?

The measuring force should be adjustable within (0-0.25) N.

What are the limits for amplifier conversion error?

Adjacent bands should not exceed 3%, and any band relative to the calibration band should not exceed 5%.