M-340Ir Compact Io/Ior Clamp Leakage Current Tester Product Leaflet

The M-340Ir is a compact, high-accuracy, high-resolution leakage current tester that is suitable for field applications. It has a voltage test lead with a fuse and can be used with a variety of large-diameter leakage current converters to expand the range of use.


M-340Ir Compact Io/Ior Clamp Leakage Current Tester Product Leaflet - cover page
Category
multimeter
Document type
Product Leaflet
Language
Chinese
Pages
2
File format
PDF
File size
99 KB
Published
16 November, 2023
Updated
16 November, 2023
MD5 checksum
E615D3637B26E3661F995913F191D173

About this manual

The M-340Ir is a compact Io/Ior clamp leakage current tester suited to field use, with high precision and high resolution for both Io and Ior down to 0.001 mA. It has fuse-protected voltage test leads and can be used with various large-diameter leakage converters to expand its application range.

The leaflet lists electrical and mechanical specifications, included accessories, optional LAD-800 and LAD-1100 leakage converters, and explains that leakage current Io consists of effective leakage current Ior and capacitive current Ioc. This manual is written in Chinese.

What's inside

  • Specifications
  • Optional accessories
  • Leakage current composition diagram
  • Ior issue
  • Test examples
  • Included accessories

Specifications

Test functionsEffective leakage current Ior and leakage current Io
Ior test circuitSingle-phase 2-wire/3-wire; three-phase 3-wire delta
Ior resolution at 10 mA range0.001 mA
Ior accuracy at 10 mA range±1.0%rdg ±10dgt
Ior resolution at 100 mA range0.01 mA
Ior accuracy at 100 mA range±1.2%rdg ±10dgt
Io resolution at 10 mA range0.001 mA
Io accuracy at 10 mA range±1.0%rdg ±10dgt
Io resolution at 100 mA range0.01 mA
Io resolution at 1000 mA range0.1 mA
Input voltage measurement pointsSingle-phase V-N; three-phase R-T
Fuse-protected test leads3 m
Data hold[DH] indicator lights and display data is held
Sampling speed2 times/second
Clamp opening diameterφ40 mm
Test circuit voltageAC 500 V or less
Operating temperature and humidity0 to 50°C, below 80% RH, non-condensing
Power supplyUM-4 ×3
Dimensions and weight200 × 44 × 24 mm, about 210 g
Included accessoriesCarrying case, UM-4 ×3, leads, instruction manual
Optional leakage convertersLAD-800, LAD-1100
LAD-800 test rangeAC 1000 A
LAD-800 CT ratio10:1, 100:1
LAD-800 CT aperture800 mm × 74 mm
LAD-800 rated voltageAC 600 V or less
LAD-800 weightabout 500 g
LAD-1100 low rangeAC 3000 mA to 30 A
LAD-1100 high rangeAC 30 A to 3000 A
LAD-1100 CT ratio10:1, 100:1
LAD-1100 CT aperture128 mm × 108 mm
LAD-1100 rated voltageAC 600 V or less
LAD-1100 weightabout 1800 g
Leakage current relationIo = Ior + Ioc

Frequently asked questions

What does the M-340Ir measure?

It measures effective leakage current Ior and leakage current Io.

What is the clamp opening diameter?

The clamp opening diameter is φ40 mm.

Which optional leakage converters can be used?

The optional leakage converters are LAD-800 and LAD-1100.

Why is Ior used to judge insulation resistance deterioration?

Insulation resistance deterioration should be judged using effective leakage current Ior.

Download PDF (99 KB)