Hitachi Magnetic Resonance Imaging System ECHELON Smart (for IRCP system) Technical Manual
This document serves as a technical manual for the Hitachi ECHELON Smart MR Imaging System, specifically designed for IRCP system integration. It is intended for service members and maintenance managers, providing essential information on system composition, technical specifications, and operational precautions.
The manual details strict prohibitions regarding installation, adjustment, maintenance, and inspection within the MRI room, emphasizing safety measures against strong magnetic fields and non-ionizing radiation. It also outlines electromagnetic compatibility (EMC) requirements, including immunity test levels and guidance for electromagnetic environments, along with a specific list of approved connecting cables to ensure optimal performance.
| Rated supply voltage | 380/400/415 V 460/480 V |
|---|---|
| Rated supply frequency | 50 Hz 60 Hz |
| Rated input | 75kVA(short time) 29kVA(continuous) |
| Protection against electric shock | CLASS I ME EQUIPMENT |
| Mode of operation | Continuous operation |
| Ambient temperature | - 15 °C to + 55 °C |
This manual is intended for service members and maintenance managers.
Servicing engineers with magnetic substances like artificial joints or pacemakers are strictly prohibited from entering the MRI room.
The MRI system must be used in a shielded location with a minimum RF shielding effectiveness of 20MHz to 80MHz, 80dB.