Hitachi ECHELON Smart MR Imaging System Technical Manual

Hitachi Magnetic Resonance Imaging System ECHELON Smart (for IRCP system) Technical Manual


Hitachi ECHELON Smart MR Imaging System Technical Manual - cover page
Brand
Hitachi
Category
Other
Document type
Service Manual
Language
English
Pages
27
File format
PDF
File size
1.4 MB
Published
16 November, 2023
Updated
16 November, 2023
MD5 checksum
51BA78D652A4A33D11824B37311B17B3

About this manual

This document serves as a technical manual for the Hitachi ECHELON Smart MR Imaging System, specifically designed for IRCP system integration. It is intended for service members and maintenance managers, providing essential information on system composition, technical specifications, and operational precautions.

The manual details strict prohibitions regarding installation, adjustment, maintenance, and inspection within the MRI room, emphasizing safety measures against strong magnetic fields and non-ionizing radiation. It also outlines electromagnetic compatibility (EMC) requirements, including immunity test levels and guidance for electromagnetic environments, along with a specific list of approved connecting cables to ensure optimal performance.

What's inside

  • System Composition
  • Technical Specification
  • Strict Prohibitions
  • Electromagnetic Compatibility
  • Cable List

Specifications

Rated supply voltage380/400/415 V 460/480 V
Rated supply frequency50 Hz 60 Hz
Rated input75kVA(short time) 29kVA(continuous)
Protection against electric shockCLASS I ME EQUIPMENT
Mode of operationContinuous operation
Ambient temperature- 15 °C to + 55 °C

Frequently asked questions

Who is this manual intended for?

This manual is intended for service members and maintenance managers.

What are the strict prohibitions for personnel entering the MRI room?

Servicing engineers with magnetic substances like artificial joints or pacemakers are strictly prohibited from entering the MRI room.

What is the minimum RF shielding effectiveness required for the MRI system location?

The MRI system must be used in a shielded location with a minimum RF shielding effectiveness of 20MHz to 80MHz, 80dB.