This document is the Solaris XE User Manual, which introduces the features and characteristics of the Solaris XE.
Peak Scientific's Solaris XE User Manual is a reference document for end users of the Solaris XE nitrogen generator. It covers safety notices, declarations, technical specifications, unpacking, installation, air purity requirements, startup and pressure setting, service, cleaning, and troubleshooting.
The manual lists operating environment limits, inlet air pressure and quality, outlet flow and purity, electrical requirements, dimensions, and weights for the generator.
| Minimum Operating Ambient Temperature | 5°C (41°F) |
|---|---|
| Maximum Operating Ambient Temperature | 35°C (95°F) |
| Maximum Altitude | 2000 m |
| Maximum Relative Humidity | 80% Non-Condensing |
| Minimum Storage Temperature | -20°C (-4°F) |
| Maximum Storage Temperature | 60°C (140°F) |
| Inlet Air Pressure | 5.51 - 9.99 bar (80 - 145 psi) |
| Minimum Air Inlet Flow | up to 100 L/min |
| Minimum Air Inlet Quality | ISO 8573-1:2010 Class[1.4.1] |
| Maximum Output Pressure | 116 psi / 7.99 bar (variable) |
| Maximum Outlet Flow Rate | up to 35 L/min |
| Purity | 95-99.5% N2 |
| Particles | <0.01µm |
| Phthalates | NONE |
| Suspended Liquids | NONE |
| Gas Outlets | 1 |
| Start-Up Time For Purity | 5 mins |
| Voltage | 100-230v |
| Frequency | 50/60 Hz |
| Current | 0.10A |
| Input Connection | C14 Plug |
| Power Cord (Supplied) | C13 Socket to Local Connection |
| Pollution Degree | 2 |
| Insulation Category | II |
| Fuses (UK Only) | 13A BS1362 |
| General Dimensions (WxDxH) | 34.3 x 65.5 x 15.6 cm (13.5 x 25.78 x 6.14 in) |
| Generator Weight | 15.1 kg (33.2 lbs) |
| Shipping Weight | 18 kg (39.6 lbs) |
| Max Weight Bearing | 100 kg (220 lbs) |
The minimum air inlet quality is ISO 8573-1:2010 Class[1.4.1].
The start-up time for purity is 5 mins.
The maximum outlet flow rate is up to 35 L/min.
It requires two people to remove the unit from the shipping crate.