Promax PE-335 Digital Earth Resistance Tester User's Manual

The PE-335 is a digital earth resistance tester with a wide measuring range, high accuracy and simple operation.


Promax PE-335 Digital Earth Resistance Tester User's Manual - cover page
Brand
Promax
Category
Test Set
Document type
User Manual
Language
Spanish
Pages
27
File format
PDF
File size
489 KB
Published
16 November, 2023
Updated
16 November, 2023
MD5 checksum
6947AA53CD0226206AF199A8B71D5ACA

About this manual

The Promax PE-335 is a digital earth resistance tester that measures earth resistance using a three-electrode method and can also measure earth voltage. It supports a simplified two-electrode method and includes a DATA HOLD function, an LCD, and a test timer.

The manual covers features, specifications, safety rules, product layout, measuring methods and maintenance, including battery replacement, cleaning and storage. It is provided in Spanish and English. This manual is written in Spanish.

What's inside

  • General
  • Features
  • Specifications
  • Safety Rules
  • Descriptive Examples of Over-Voltage Categories
  • Product Layout
  • Measuring Methods
  • Maintenance
  • Battery Replacement
  • Cleaning and Storage

Specifications

Measurement systemConstant current injection at approx. 820 Hz, 2 mA
Earth voltage range0 to 199.9 V AC, 50/60 Hz
Maximum input voltage200 V rms
Earth resistance range 10 to 19.99 Ω, resolution 0.01 Ω
Earth resistance range 20 to 199.9 Ω, resolution 0.1 Ω
Earth resistance range 30 to 1999 Ω, resolution 1 Ω
Earth resistance accuracy±(2% reading + 2 digits) or ±0.1 Ω, whichever is greater
Earth voltage accuracy±(1% reading + 2 digits)
Response time4 to 8 seconds
DisplayLCD 3 1/2 digits (2000 counts)
Power9 V DC (6 x 1.5 V R6 batteries)
WeightApprox. 600 g including batteries
Operating temperature-10 °C to 55 °C
TimerAuto-disconnect 3 to 5 min after TEST ON/OFF

Frequently asked questions

What type of batteries does the PE-335 use?

It uses six 1.5 V R6 batteries, giving a 9 V DC supply.

What is the measurement current of the earth resistance tester?

The measurement current is approximately 2 mA, which helps avoid activating the limiter of the circuit under test.

What should I do if the +- symbol appears on the display?

Replace the batteries with new ones to avoid incorrect measurements.

How long does the PE-335 take to respond?

The response time is from 4 to 8 seconds.

Download PDF (489 KB)

More Promax Manuals