Metrohm 885 Compact Oven SC Oven Module User Manual

The 885 Compact Oven SC is a compact oven specifically designed for Karl Fischer moisture determination. Its features include:


Metrohm 885 Compact Oven SC Oven Module User Manual - cover page
Brand
Metrohm
Category
Other
Document type
User Manual
Language
Chinese
Pages
77
File format
PDF
File size
2.7 MB
Published
15 November, 2023
Updated
15 November, 2023
MD5 checksum
75E86B95D0315215D3E405CC75D7E749

About this manual

The Metrohm 885 Compact Oven SC is an oven module used when samples must be heated or solid and liquid materials thermally dried. Combined with a coulometric or volumetric titrator, it forms a water-content analysis system for samples with destructive components or moisture that is difficult to evaporate.

This Chinese-language manual covers instrument components, safety, installation, automatic processes, operation, method handling, sample series, system settings, determinations, maintenance, troubleshooting and technical data such as flow range, power connection, ambient conditions and dimensions. This manual is written in Chinese.

What's inside

  • Introduction
  • Safety
  • Instrument overview
  • Installation
  • Automatic processes
  • Operation
  • System settings
  • Performing determinations
  • Troubleshooting
  • Technical data

Specifications

Flow range10–150 mL/min
Voltage100–120 / 220–240 V
Frequency50 / 60 Hz
Power consumption200 W
Fuse2.0 ATH
Width0.28 m
Height0.45 m
Depth0.44 m
Weight12.03 kg (without accessories)
Nominal ambient range+5 to +45 °C
Maximum relative humidity80%, non-condensing
Typical tubing temperatureapprox. 50 °C

Frequently asked questions

What types of samples is the 885 Compact Oven SC intended for?

It is intended for samples containing destructive components or samples whose moisture is difficult to evaporate.

How many sample positions does the turntable have?

It has 17 sample bottle positions and 1 position for an equilibration tube.

What is the gas flow range?

The gas flow range is 10–150 mL/min.

What mains voltage does the instrument support?

It supports 100–120 / 220–240 V.

Download PDF (2.7 MB)

More Metrohm Manuals