Metrohm application note AN-V-127 for determining Fe(II) in sucrose injection with the 884 voltammeter according to USP-NF.
This application note (AN-V-127) describes the determination of Fe(II) in sucrose injection with the Metrohm 884 voltammeter according to USP-NF requirements. It explains that Fe(II) and Fe(III) give different signals at different potentials, so Fe(II) can be measured without prior separation.
The document lists method parameters including an SMDE working electrode, differential pulse mode, start potential -0.1 V and end potential -1.75 V, and peak potentials for Fe(III) and Fe(II). A determination result of 0.33% Fe(II) in sucrose injection is reported; control, data acquisition and evaluation use viva software. This manual is written in Chinese.
| Working electrode | SMDE |
|---|---|
| Reference electrode | Ag/AgCl/KCl (3 mol/L) |
| Control and evaluation software | viva |
It describes determination of Fe(II) in sucrose injection according to USP-NF requirements using the 884 voltammeter.
The reported Fe(II) content in sucrose injection was 0.33%.
The method specifies an Ag/AgCl/KCl (3 mol/L) reference electrode.
viva software is used for control, data acquisition and evaluation.