The DXW-2A Electric Gastric Lavage Apparatus is a medical device used for emergency gastric lavage in cases of poisoning, food poisoning and before surgery.
This document is the instructions for use of the Simanfeng DXW-2A electric gastric lavage apparatus (电动洗胃机), published by Shanghai Baojia Medical Device Co., Ltd. The machine is intended for use in medical units for emergency lavage in oral poisoning and food poisoning cases, and for pre-operative gastric lavage. It consists of an air pump, control circuit with software, a reversing device, gastric fluid and medication containers, and solenoid valves.
The manual describes the device's contraindications, safety warnings, working principle with a pressure feedback control system, front and rear panel controls, performance specifications, tubing installation and commissioning, the lavage working process, cleaning and disinfection, maintenance, EMC information, troubleshooting, waste disposal, operating and storage conditions, after-sales service and accessories. It also covers the oral and nasal intubation lavage modes and the fluid balance function. This manual is written in Chinese.
| Flow rate | ≥2 L/min (oral intubation); ≥1 L/min (nasal intubation) |
|---|---|
| Automatic fluid volume | Flush 250–350 ml/cycle; suction 350–450 ml/cycle |
| Positive/negative pressure setting range | 47 kPa – 67 kPa |
| Power supply | AC 220 V, 50 Hz |
| Input power | 110 VA |
| Fuse | 5×20 F750mA L/250V |
| Noise | ≤65 dB(A) |
| Net weight | 10 kg |
| Dimensions | 430 mm × 360 mm × 205 mm |
| Protection class | Class I equipment, Type B applied part, IPX0 |
| Service life | Five years (diaphragm of medication and gastric fluid containers: one year) |
Contraindicated patients include those who swallowed corrosive substances such as strong acids or alkalis, patients with upper GI ulcers or cancer, patients with intubation contraindications such as esophageal obstruction or varices, comatose patients, patients at risk of bleeding or gastric perforation, and infants or young children.
When fluid intake exceeds fluid output, pressing the fluid balance key reduces the intake volume by about 150–300 ml per cycle; pressing it 2–3 times before the end of lavage helps clear residual fluid from the stomach.
In high-altitude areas always use the oral intubation position, and in plain areas use it when lavaging with a gastric tube with an inner diameter larger than F24; tubes with an inner diameter of F24 or less use the nasal intubation position.
No, the lavage machine is not suitable for use in an MRI environment and should not be used in other strong magnetic field environments.