The force辰technology electronic precision balance adopts high precision, high sensitivity and high stability electromagnetic force balance sensor, which makes the measurement result more accurate, the response speed faster and the data stability higher. The touch feedback key is adopted to ensure accurate and effective operation. The balance is designed and manufactured with a fully transparent sealed cover to avoid the influence of air flow on the weighing result. It can be widely used in industry, agriculture, commerce, schools, scientific research and other units to quickly determine the quality and quantity of objects.
The Lichen Keji SJT Series is a line of electronic precision balances that uses an electromagnetic force compensation sensor, an LCD display and a fully transparent draft shield. The manual lists standard models LC-SJT603, LC-SJT803, LC-SJT1003 and LC-SJT2003 with maximum capacities of 600 g, 800 g, 1000 g and 2000 g, plus internal-calibration C models LC-SJT603C, LC-SJT803C and LC-SJT1003C.
The document explains leveling, warm-up, external or internal calibration, weighing functions, RS-232C output, maintenance and troubleshooting. It states that the balance is suitable for industrial, agricultural, commercial, school and research use. This manual is written in Chinese.
| Models | LC-SJT603, LC-SJT803, LC-SJT1003, LC-SJT2003 |
|---|---|
| Input power | AC110V-240V |
| Maximum capacity | 600g, 800g, 1000g, 2000g |
| Minimum capacity | 0.001g |
| Display | LCD liquid crystal display |
| Accuracy class | Class II |
| Readability | 0.001g |
| Repeatability error | ±1mg, ±3mg |
| Linearity error | ±2mg, ±4mg |
| Stabilization time | 1s, 3s |
| Pan size | Φ90mm |
| Calibration | External weight calibration |
| Weighing units | g/oz/ct/mg/custom |
| RS232 interface | Yes |
| Draft shield dimensions | 180×195×215mm |
| Net weight | 5.5Kg |
| Product dimensions (L×W×H) | 310×212×350mm |
| Internal calibration models | LC-SJT603C, LC-SJT803C, LC-SJT1003C |
| Internal calibration | Fully automatic internal calibration |
The standard models LC-SJT603, LC-SJT803, LC-SJT1003 and LC-SJT2003 have maximum capacities of 600 g, 800 g, 1000 g and 2000 g.
Yes. The manual states that the instrument should be preheated for at least 60 minutes before operation to ensure accuracy.
Calibration is required before the first weighing, after a long power-off or power failure, after changing the operating environment and periodically during weighing.
An RS-232C interface is installed at the rear of the balance, and a USB interface can be customized.