The pen pH meter is a portable instrument that can display pH value and temperature value simultaneously. It is suitable for water quality, pH value measurement in water supply monitoring, aquaculture, environmental protection, printing and dyeing, electroplating, beverage industry and scientific research units.
The Lichen Technology pH-10/100/100A/100B pen pH meters are designed for convenient carrying and field or outdoor operation, and can display pH and temperature simultaneously. They are suitable for tap water source monitoring, aquaculture, environmental protection, dyeing, electroplating, beverage industry and research units for water quality and pH determination.
The meters have an LCD display, automatic temperature compensation, temperature unit conversion, data hold, waterproof construction with a buoyancy device, and a replaceable electrode head; pH-100A adds a backlit display and pH-100B supports an external long-line composite electrode. The manual covers precautions, product structure, features, parameters, operating instructions, maintenance, warranty, unpacking inspection and packing list. This manual is written in Chinese.
| pH measurement range | 0.1-14.0 pH (pH-10); 0.01-14.00 pH (pH-100/100A/100B) |
|---|---|
| Temperature measurement range | 0-50 °C (32-122 °F) |
| pH resolution | 0.1 pH (pH-10); 0.01 pH (pH-100/100A/100B) |
| pH accuracy | ±0.1 pH (pH-10); ±0.02 pH (pH-100/100A/100B) |
| pH calibration | 25 °C three-point calibration: 4.0, 6.9, 9.2 (pH-10); 4.01, 6.86, 9.18 (pH-100/100A/100B) |
| Operating temperature | 0-50 °C (32-122 °F) |
| Power | 1.5 V LR44 ×4 (pH-10/100/100B); 3 V LR2032 ×4 (pH-100A) |
| Dimensions | Φ40×185 mm (pH-10/100/100A); Φ40×180 mm (pH-100B) |
Prepare standard buffer solution, turn the meter on, insert the electrode, and press CAL for about three seconds; the meter automatically recognizes the buffer value, then shows SA and END when calibration is complete.
With the meter off, hold CAL and press ON/OFF; when °C or °F appears, release CAL, press CAL to select the unit, then press HOLD to save.
It means the battery voltage is insufficient, so the batteries should be replaced soon; do not mix old and new batteries.
After each test, rinse it clean with purified or distilled water; if the meter will not be used for a long time, remove the batteries.