Lichen Keji SZF Series Crude Fat Analyzer Instruction Manual

SZF series crude fat analyzer is a device to measure fat content by weight measurement method. It can adjust the heating temperature according to the boiling point of the reagent and the ambient temperature. It is an ideal device for fat measurement in food, fat, and feed industries.


Lichen Keji SZF Series Crude Fat Analyzer Instruction Manual - cover page
Brand
力辰科技
Category
Analyzer
Document type
User Manual
Language
Chinese
Pages
14
File format
PDF
File size
633 KB
Published
30 October, 2023
Updated
30 October, 2023
MD5 checksum
CB725F9DFD7DAD13EFFCBFA9CD637390

About this manual

The Lichen Keji SZF Series crude fat analyzer determines fat content by the Soxhlet extraction principle and a gravimetric method: fat is dissolved in an organic solvent, separated by extraction, then dried, weighed and calculated. The instrument is described as consisting of heating/immersion extraction, solvent recovery and cooling sections, for fat determination in food, fat and feed industries.

The manual covers the SZF-06A, SZF-06B, SZF-06C and SZF-06 models, their structure, features and parameters, plus equipment installation, use, maintenance and fault analysis. It includes warnings about petroleum ether use in a fume hood, cooling water connections and water bath operation. This manual is written in Chinese.

What's inside

  • Overview
  • Structure
  • Product Features
  • Product Parameters
  • Equipment Installation
  • Equipment Use
  • Maintenance
  • Fault Analysis
  • Warranty Statement
  • Packing List

Specifications

Input powerAC 220 V / 50 Hz
Rated power800 W / 500 W (model-dependent)
Measurement rangeOil content 0.5%–60% in grain, feed, oilseed and fat products
Sample weight0.5–15 g (routine 2–5 g)
Samples per batch6
Heating methodWater bath heating; electric hot plate heating
DisplayLED digital tube
Temperature control rangeRoom temperature to 100 °C (water bath); room temperature +5 °C to 300 °C (hot plate)
Solvent recovery rate≥80%

Frequently asked questions

What principle is used to determine fat content?

The analyzer uses the Soxhlet extraction principle and a gravimetric method: fat is dissolved in an organic solvent, separated by extraction, dried, weighed and calculated.

How much sample is used?

The sample weight range is 0.5–15 g, with a routine amount of 2–5 g.

Where must the analyzer be operated when using petroleum ether?

Because petroleum ether is highly volatile, operation must be carried out in a fume hood.