Time Electronics 1006 / 1007 DC Millivolt Source and Potentiometer User Manual

1006 DC Millivolt Source and 1007 DC Millivolt Potentiometer/Source User Manual


Time Electronics 1006 / 1007 DC Millivolt Source and Potentiometer User Manual - cover page
Brand
Time Electronics
Category
Signal Generators
Document type
User Manual
Language
English
Pages
15
File format
PDF
File size
1 MB
Published
10 November, 2023
Updated
10 November, 2023
MD5 checksum
AC45915E8DC20CD3BE4E775E2713DC91

About this manual

This is the Time Electronics user manual (version 1.2) for the 1006 DC Millivolt Source and the 1007 DC Millivolt Potentiometer and Source. It introduces both instruments, lists their features and specifications, describes the front panel controls, and gives step-by-step operating instructions. A section on servicing and calibration covers the internal preset controls and adjustment, followed by warranty and servicing information.

The 1006 is a millivolt source with three output ranges from 1 µV to 1 V and 0.02 % accuracy, powered by 6 AA batteries and protected against short circuit and overload. The 1007 adds a microvolt null balance display so it can also be used for potentiometric voltage measurement. The manual explains battery insertion, source operation on both models, and the zero setting and measuring procedures for the 1007 null detector.

What's inside

  • Introduction
  • 1006 DC Millivolt Source features
  • 1007 DC Millivolt Potentiometer and Source features
  • Specifications
  • 1006 Controls
  • 1007 Controls
  • Battery Insertion
  • 1006/1007 mV Source Operation
  • 1007 mV Potentiometric Null Operation
  • Calibration and Adjustment

Specifications

Output ranges0 to 999.9 mV in 3 ranges: 0 to 999.9 mV in 0.1 mV steps, 0 to 99.99 mV in 10 µV steps, 0 to 9.999 mV in 1 µV steps
Accuracy± (0.02 % of setting + 0.02 % of range + 1 µV)
Output resistanceLess than 0.2 Ω on 1 V and 100 mV ranges; 1 Ω on 10 mV range
Maximum output current20 mA on 999.9 mV and 99.99 mV ranges
Output voltage stabilityLess than 60 ppm/°C; less than 100 ppm per 3 months (non cumulative)
Output noise levelLess than 30 ppm of full scale
Operating temperature-10 °C to +60 °C
Power supply6 AA batteries; optional 9529 rechargeable battery pack (6 cells and mains charger)
Battery life100 hours typical
Null balance display (1007 only)Maximum sensitivity ±20 µV fsd, minimum sensitivity ±200 mV fsd, input resistance greater than 1 MΩ at balance
Dimensions200 x 75 x 110 mm (215 x 100 x 120 mm incl. protective cover)
Weight0.75 kg (1.2 kg incl. protective cover)

Frequently asked questions

How do I insert the batteries in the 1006 / 1007?

Remove the two black covers on top of the instrument by pressing in and turning them 90° anti-clockwise, then insert three AA batteries in each tube with the positive terminal up. NiMH rechargeable batteries must be used if the optional battery charger is used.

What do the colors of the battery level indicator mean?

Green means the unit is ready for use; red means the batteries should be changed or recharged.

How do I zero the 1007 null detector before measuring?

Set the function switch to "Zero", select the range matching the voltage to be measured, set the sensitivity dial to maximum, and adjust the zero control dial for a zero reading on the null display.

What is the difference between the 1006 and 1007 output polarity switch in the off position?

On the 1006 the off position disconnects the output and provides a short circuit on the output terminals, while on the 1007 it provides an open circuit to prevent accidental shorting of the voltage under test.