1006 DC Millivolt Source and 1007 DC Millivolt Potentiometer/Source User Manual
This is the Time Electronics user manual (version 1.2) for the 1006 DC Millivolt Source and the 1007 DC Millivolt Potentiometer and Source. It introduces both instruments, lists their features and specifications, describes the front panel controls, and gives step-by-step operating instructions. A section on servicing and calibration covers the internal preset controls and adjustment, followed by warranty and servicing information.
The 1006 is a millivolt source with three output ranges from 1 µV to 1 V and 0.02 % accuracy, powered by 6 AA batteries and protected against short circuit and overload. The 1007 adds a microvolt null balance display so it can also be used for potentiometric voltage measurement. The manual explains battery insertion, source operation on both models, and the zero setting and measuring procedures for the 1007 null detector.
| Output ranges | 0 to 999.9 mV in 3 ranges: 0 to 999.9 mV in 0.1 mV steps, 0 to 99.99 mV in 10 µV steps, 0 to 9.999 mV in 1 µV steps |
|---|---|
| Accuracy | ± (0.02 % of setting + 0.02 % of range + 1 µV) |
| Output resistance | Less than 0.2 Ω on 1 V and 100 mV ranges; 1 Ω on 10 mV range |
| Maximum output current | 20 mA on 999.9 mV and 99.99 mV ranges |
| Output voltage stability | Less than 60 ppm/°C; less than 100 ppm per 3 months (non cumulative) |
| Output noise level | Less than 30 ppm of full scale |
| Operating temperature | -10 °C to +60 °C |
| Power supply | 6 AA batteries; optional 9529 rechargeable battery pack (6 cells and mains charger) |
| Battery life | 100 hours typical |
| Null balance display (1007 only) | Maximum sensitivity ±20 µV fsd, minimum sensitivity ±200 mV fsd, input resistance greater than 1 MΩ at balance |
| Dimensions | 200 x 75 x 110 mm (215 x 100 x 120 mm incl. protective cover) |
| Weight | 0.75 kg (1.2 kg incl. protective cover) |
Remove the two black covers on top of the instrument by pressing in and turning them 90° anti-clockwise, then insert three AA batteries in each tube with the positive terminal up. NiMH rechargeable batteries must be used if the optional battery charger is used.
Green means the unit is ready for use; red means the batteries should be changed or recharged.
Set the function switch to "Zero", select the range matching the voltage to be measured, set the sensitivity dial to maximum, and adjust the zero control dial for a zero reading on the null display.
On the 1006 the off position disconnects the output and provides a short circuit on the output terminals, while on the 1007 it provides an open circuit to prevent accidental shorting of the voltage under test.