Time Electronics 1051 Resistance Decade Box User Manual

The 1051 resistance decade box is a precision resistance decade box designed for general laboratory work including platinum resistance thermometry.


Time Electronics 1051 Resistance Decade Box User Manual - cover page
Brand
Time Electronics
Category
Calibration Components
Document type
User Manual
Language
English
Pages
13
File format
PDF
File size
1.3 MB
Published
02 November, 2023
Updated
02 November, 2023
MD5 checksum
32B36CD2DDB5D49160DB1D5606D47493

About this manual

The Time Electronics 1051 is a precision resistance decade box designed for general laboratory work, including platinum resistance thermometry. It uses high-stability metal film resistors, an 8-digit thumbwheel switch, color-coded digits, safety terminals, and a removable protective rubber cover.

The manual provides specifications, safety precautions, front panel control descriptions, operating instructions, a resistance setting table, applications such as Pt100 resistance simulation, and maintenance and repair information including overload considerations and spare parts.

What's inside

  • Introduction
  • Specifications
  • Safety Precautions
  • Front Panel Controls and Connections
  • Operating Instructions
  • Resistance Setting Table
  • Applications
  • Maintenance and Repair
  • Warranty and Servicing

Specifications

Resistance range0.01 Ω to 1 MΩ in 0.01 Ω steps
Residual resistanceLess than 90 mΩ (< 11.25 mΩ / decade)
Power rating1 watt per resistor
Voltage ratingMaximum 250 V DC, 176 V AC RMS
Current ratingMaximum 1.0 A DC, 0.707 A AC RMS
Temperature coefficient50 ppm/°C
Connections2 x 4 mm active safety terminals
DimensionsW200 x H75 x D110 mm
Weight0.6 kg (1.0 kg including protective cover)

Frequently asked questions

What is the resistance range of the 1051?

It covers 0.01 Ω to 1 MΩ in 0.01 Ω steps.

How do I set 138.51 Ω for a Pt100 at 100 °C?

Set the 100 Ω decade to 1, the 10 Ω decade to 3, the 1 Ω decade to 8, the 0.1 Ω decade to 5, and the 0.01 Ω decade to 1; set the higher decades to 0.

Should residual resistance be subtracted for precision measurements?

Yes. For precision measurements or recalibration, the residual value must be subtracted from all measurements.

What is the maximum voltage that can be applied?

The maximum voltage is 250 V DC or 176 V AC RMS.