The RS-PH-N01-3 integrated PH transmitter is a device that measures the PH value (hydrogen ion concentration index, acidity and alkalinity) of a solution, with automatic temperature compensation function, and automatic temperature compensation and manual temperature compensation can be switched at will.
This is the V1.2 user manual for the Renke RS-PH-N01-3 integrated pH transmitter, a device that measures the pH value of a solution with 0~14.00 pH range and 0.01 pH resolution, plus temperature from 0~60 °C. It has an integrated design, IP68 protection, automatic or manual temperature compensation, and an RS485 interface using the ModBus-RTU protocol; it is intended for non-corrosive weak acid and weak alkali environments such as industrial and domestic wastewater, agriculture and aquaculture.
The manual lists the technical parameters, model selection codes and package contents, then explains dimensions and mounting (immersion and pipe installation), wiring colors, parameter configuration with the 485 configuration software, two-point electrode calibration with standard buffer solutions, and the ModBus communication parameters, register addresses and example frames. It ends with precautions, maintenance notes, contact details and a document history. This manual is written in Chinese.
| Power supply | DC 7~30V |
|---|---|
| pH measurement range / resolution | 0~14.00 pH / 0.01 pH |
| pH measurement error | ±0.15 pH |
| Repeatability error | ±0.02 pH |
| Temperature measurement range / resolution | 0~60 ℃ / 0.1 ℃ |
| Temperature measurement error | ±0.5 ℃ |
| Communication interface | RS485, ModBus-RTU, default baud rate 4800 (1200~115200 selectable) |
| Ingress protection | IP68 |
| Pressure resistance | 0.6 MPa |
| Cable length | 5 m default (other lengths customizable) |
| Electrode service life | 6~12 months |
| Operating ambient temperature | 0~60 ℃ |
It measures 0~14.00 pH with 0.01 pH resolution and 0~60 ℃ with 0.1 ℃ resolution.
The factory default baud rate is 4800 bit/s and the default address is 0x01.
Use 4.01 and 6.86 for acidic solutions, 6.86 and 9.18 for alkaline solutions, or 4.01 and 9.18 for general use.
Brown is power positive, black is power negative, yellow is 485-A and blue is 485-B.