Renke RS-PH-N01-3 Integrated pH Transmitter User Manual

The RS-PH-N01-3 integrated PH transmitter is a device that measures the PH value (hydrogen ion concentration index, acidity and alkalinity) of a solution, with automatic temperature compensation function, and automatic temperature compensation and manual temperature compensation can be switched at will.


Renke RS-PH-N01-3 Integrated pH Transmitter User Manual - cover page
Brand
Renke
Category
sensor
Document type
User Manual
Language
Chinese
Pages
13
File format
PDF
File size
516 KB
Published
07 November, 2023
Updated
07 November, 2023
MD5 checksum
ADB8107C7AA8F2371FA8EBC27C0B9F77

About this manual

This is the V1.2 user manual for the Renke RS-PH-N01-3 integrated pH transmitter, a device that measures the pH value of a solution with 0~14.00 pH range and 0.01 pH resolution, plus temperature from 0~60 °C. It has an integrated design, IP68 protection, automatic or manual temperature compensation, and an RS485 interface using the ModBus-RTU protocol; it is intended for non-corrosive weak acid and weak alkali environments such as industrial and domestic wastewater, agriculture and aquaculture.

The manual lists the technical parameters, model selection codes and package contents, then explains dimensions and mounting (immersion and pipe installation), wiring colors, parameter configuration with the 485 configuration software, two-point electrode calibration with standard buffer solutions, and the ModBus communication parameters, register addresses and example frames. It ends with precautions, maintenance notes, contact details and a document history. This manual is written in Chinese.

What's inside

  • 产品介绍 (Product Introduction)
  • 功能特点 (Features)
  • 设备技术参数 (Technical Parameters)
  • 产品选型 (Product Selection)
  • 产品清单 (Package Contents)
  • 设备尺寸及安装 (Dimensions and Installation)
  • 接线说明 (Wiring Instructions)
  • 参数配置说明 (Parameter Configuration)
  • 电极标定说明 (Electrode Calibration)
  • ModBus 通信及寄存器详解 (ModBus Communication and Registers)

Specifications

Power supplyDC 7~30V
pH measurement range / resolution0~14.00 pH / 0.01 pH
pH measurement error±0.15 pH
Repeatability error±0.02 pH
Temperature measurement range / resolution0~60 ℃ / 0.1 ℃
Temperature measurement error±0.5 ℃
Communication interfaceRS485, ModBus-RTU, default baud rate 4800 (1200~115200 selectable)
Ingress protectionIP68
Pressure resistance0.6 MPa
Cable length5 m default (other lengths customizable)
Electrode service life6~12 months
Operating ambient temperature0~60 ℃

Frequently asked questions

What is the pH and temperature measurement range of the RS-PH-N01-3?

It measures 0~14.00 pH with 0.01 pH resolution and 0~60 ℃ with 0.1 ℃ resolution.

What are the default ModBus baud rate and device address?

The factory default baud rate is 4800 bit/s and the default address is 0x01.

Which standard buffer solutions are used for electrode calibration?

Use 4.01 and 6.86 for acidic solutions, 6.86 and 9.18 for alkaline solutions, or 4.01 and 9.18 for general use.

How is the transmitter wired?

Brown is power positive, black is power negative, yellow is 485-A and blue is 485-B.