This product is a radar level meter that uses a special related demodulation technology to accurately identify the time interval between the transmitted pulse and the received pulse, and then further calculate the distance from the antenna to the measured medium surface. Even in complex working conditions, with the latest microprocessor technology and debugging software, the echo of the material can be accurately analyzed to obtain accurate measurement results.
This Chinese user manual covers the Sinomeasure 80 series low-frequency radar level meters (models 801, 802, 803, 803T, 804 and 805). It explains the microwave pulse measuring principle, the distance and level calculation, and the applications of the 6 GHz series for contactless continuous level measurement of liquids, slurries, granules and bulk solids.
The manual describes the individual model versions and their antennas, then walks through installation (tank shape, mounting position, stilling well, nozzle height), electrical connection and supply, commissioning via display/keys, PC or HART handheld programmer, housing and horn dimensions, technical parameters, maintenance notes, and warranty and after-sales service. Document number U-SIN-RD80-CN2, second edition, July 2021. This manual is written in Chinese.
| Frequency | 6 GHz |
|---|---|
| Measuring range (801 / 802) | 0–20 m |
| Measuring range (803) | 0–35 m |
| Measurement accuracy | ±10 mm at 5 m range; ±20 mm at 20 m range |
| Output signal | (4–20) mA / HART |
| Resolution | 1.6 µA |
| Supply voltage, two-wire | 24 VDC |
| Supply voltage, four-wire | (198–242) VAC |
| Power consumption | max. 22.5 mA |
| Protection class | IP66/67 |
| Maximum tank pressure | Max. 4 MPa |
| Display resolution | 1 mm |
| Process temperature (801) | (-40~130) °C |
| Relative humidity | <95% |
Both the 801 and 802 have a measuring range of 0 to 20 m. The 803 reaches 0 to 35 m.
The output is (4–20) mA with HART, and the meter can be supplied as a two-wire version with 24 VDC or a four-wire version with 24 VDC or 220 VAC.
No. The manual states the product must not be used in explosion-proof (hazardous) locations.
Accuracy is ±10 mm at a 5 m range and ±20 mm at a 20 m range for the 801 and 802; the 803 is ±15 mm at 20 m and ±20 mm at 35 m.