This manual covers the operation, installation, maintenance, and Modbus RTU communication protocol for the Sinomeasure U-SIN-pH-8001 pH digital electrode.
The Sinomeasure U-SIN-pH-8001 is a digital pH and ORP electrode designed for aquaculture applications. It features a compact structure, isolated power design for stability, and supports automatic or manual temperature compensation. The device connects via an RS485 interface using the standard Modbus RTU protocol, allowing remote configuration and data acquisition from a host computer.
This manual provides instructions on verifying package contents, establishing electrical connections, and performing daily maintenance such as cleaning and storage. It also details the communication protocol, including how to query the device address, execute calibration commands for various buffer solutions (USA/NIST/Custom), and restore factory settings. This manual is written in Chinese.
| Measurement Range | 0.00~14.00 pH |
|---|---|
| Resolution | 0.01pH |
| Temperature Range | 0~60 ℃ |
| Output Interface | RS485 |
| Communication Protocol | Modbus-RTU |
| Address Range | 1-255 |
| Power Supply | 6~30 VDC |
| Power Consumption | 30mA@12VDC |
It is recommended to replace the electrode every 12 months.
Use a rubber cap and wet sponge to keep the electrode moist.
Yes, the protocol supports modifying the ID from 1 to 255.
The device supports USA, NIST, and custom pH calibration solutions.