This file is the 904 Titrando manual, which introduces the assembly, installation, connection, operation and maintenance of the instrument.
The Metrohm 904 Titrando is the core component of the modular Titrando system. It can be operated with a Touch Control with touch screen or with a PC running the appropriate software. The instrument has an internal dosing unit for the exchange unit, four MSB interfaces, and two USB interfaces.
Depending on the instrument model, it has one or two measurement interfaces. Each measurement input supports electrodes for potentiometry, a separate reference electrode, a temperature sensor, a polarized electrode, or iConnect. The manual covers installation, connection of controllers and accessories, titration cup setup, operation and maintenance, troubleshooting, remote interface pin assignments, technical data, and accessories. This manual is written in Chinese.
| Power supply voltage | 100-240 V |
|---|---|
| Power frequency | 50-60 Hz |
| Power consumption | Maximum 45 W |
| Operating temperature | +5 to +45 C |
| Operating humidity | Maximum 80% relative humidity, non-condensing |
| Storage temperature | +5 to +45 C |
| Width | 142 mm |
| Height | 227 mm |
| Depth | 231 mm |
| Weight (model 2.904.0010) | 2.9 kg without accessories |
| Weight (model 2.904.0020) | 3.1 kg without accessories |
| Housing material | PBT |
| USB downstream ports | 2 (Type A), 500 mA each |
| USB controller port | USB upstream port with additional power (Mini-DIN) |
| MSB interfaces | 4 |
| External dosing units | Up to 3 external Dosimat or Dosino dosing units (MSB 2 to MSB 4) |
| Stirrers | Up to 4 |
| Remote Box units | Up to 4 |
| Measurement interfaces | One or two, depending on instrument model |
It has two USB interfaces for connecting peripherals such as printers, keyboards, barcode readers, or other control devices.
It has four MSB interfaces for connecting dosing units, stirrers, titration stands, and Remote Box units.
The nominal operating range is +5 to +45 C.
It has one or two measurement interfaces, depending on the instrument model.