This manual introduces the functions and features of the R1L-B/R1L-BR microohmmeter and bond tester, including instrument description, operating instructions, and maintenance
This manual provides detailed instructions for operating, maintaining, and troubleshooting the TEGAM R1L-B/R1L-BR microohmmeter and bond tester. It includes sections on instrument description, preparation for use, operating instructions, principles of operation, maintenance, and service information. The R1L-B model has a bench-top case, while the R1L-BR has a rugged portable case. Both models use a four-terminal Kelvin technique for accurate resistance measurements and feature a 3½-digit LCD display.
The manual also covers safety precautions, battery installation, calibration, warranty terms, and repair procedures. It includes tables for specifications, parts lists, and troubleshooting symptoms. The device is powered by rechargeable NiCad batteries and can operate in a wide temperature and humidity range.
| Full Scale Ranges | 2 mΩ to 20 Ω |
|---|---|
| Display | 3½ digit liquid crystal display |
| Accuracy | ± (0.25% of reading + 1 count) |
| Battery Type | Three rechargeable NiCad ‘C’ cells (2800 mAh - 3500 mAh) |
| Operating Temperature | 0 °C to 50 °C |
| Relative Humidity | 75% RH non-condensing |
| Altitude | Up to 3050 m |
Accuracy on all ranges: ± (0.25% of reading + 1 count).
It is a digital Microohmmeter and Bond Tester that uses a four-terminal Kelvin technique to eliminate lead resistance errors.
The R1L-BR measures 127 mm (5") x 254 mm (10") x 228 mm (9") and weighs 3 kg (6.7 lb).
TEGAM warrants the product to be free from defects in material and workmanship for a period of three years from the date of shipment.