Gilson GX-281 Liquid Handler User's Guide introduces the liquid handler's safety, description, usage, maintenance and more
This guide provides step-by-step instructions for installing, operating, and maintaining the Gilson GX-281 Liquid Handler. It includes safety information, setup procedures for components like probes and solvents, operational controls, and troubleshooting tips. Maintenance sections cover cleaning, part replacement, and transport guidelines. Technical specifications and GSIOC configuration details are also included for advanced users.
| Arm Speed (X) | 0.01 to 600.00 mm/sec |
|---|---|
| Dimensions (W x D x H) | 88.9 x 76.2 x 96.5 cm (35 x 30 x 38 in) |
| Display | Two lines of 40 alphanumeric characters; vacuum fluorescent display |
| Environmental Conditions | Indoor use, up to 2000 m altitude, 5°–40°C temperature, 75–105 kPa air pressure, 80% max humidity at 31°C decreasing to 50% at 40°C |
| Fuse | 5 x 20 mm “T” type 5.0A (one for 100–120V; two for 220–240V) |
The arm speed ranges are 0.01 to 600.00 mm/sec in X and Y dimensions, and 0.01 to 125.00 mm/sec in Z dimension.
The dimensions are 88.9 x 76.2 x 96.5 cm (35 x 30 x 38 in).
The display is a two-line vacuum fluorescent display with 40 alphanumeric characters per line.
The GX-281 is for indoor use, with an altitude limit of 2000 m, temperature range of 5°–40°C, air pressure of 75–105 kPa, and humidity up to 80% at 31°C decreasing to 50% at 40°C.