Hanna Instruments HI99121 Waterproof Soil pH and Temperature Meter Instruction Manual

The document is the instruction manual for Hanna Instruments' HI99121 soil pH and temperature meter


Hanna Instruments HI99121 Waterproof Soil pH and Temperature Meter Instruction Manual - cover page
Brand
Hanna Instruments
Category
Other Electronics
Document type
User Manual
Language
English
Pages
24
File format
PDF
File size
2.8 MB
Published
26 October, 2023
Updated
26 October, 2023
Model
HI99121
MD5 checksum
25064F204227BEEF379548A72C285D81

About this manual

The HI99121 is a portable, waterproof pH and temperature meter supplied with the HI12923 pH/temperature probe. It is designed for direct pH measurement in soil or soil slurries, with a conical probe tip, built-in temperature sensor, dual-line LCD, and automatic pH calibration at one or two points.

The manual covers preliminary examination, main features, specifications, display description, operation, meter setup, soil and substrate measurement, battery replacement, accessories, and electrode maintenance. It also explains calibration with standard or NIST buffer sets and use of the supplied HI721319 auger to loosen soil before measurement.

What's inside

  • Preliminary examination
  • General description and intended use
  • Main features
  • Specifications
  • Display description
  • Operational guide
  • Meter setup
  • Soil pH
  • Battery replacement
  • Electrode maintenance

Specifications

pH range-2.00 to 16.00 pH / -2.0 to 16.0 pH
pH resolution0.01 pH / 0.1 pH
pH accuracy±0.02 pH / ±0.1 pH
Temperature compensationAutomatic, -5.0 to 105.0 °C / 23.0 to 221.0 °F
pH calibrationAutomatic, 1 or 2 point; standard or NIST buffer set
Included probeHI12923 preamplified pH and temperature probe with conical tip, DIN connector and 1 m cable
Battery type and life1.5V AAA (3 pcs.), approx. 1400 hours of continuous use
Operating environment0 to 50 °C (32 to 122 °F), RH max. 100%
Dimensions154 x 63 x 30 mm (6.1 x 2.5 x 1.2 in.)
Weight with batteries196 g (6.91 oz.)
Ingress protectionIP67

Frequently asked questions

What buffers does the HI99121 use for pH calibration?

It uses automatic one- or two-point calibration with a selectable standard buffer set (4.01, 7.01, 10.01) or NIST buffer set (4.01, 6.86, 9.18).

How long does the HI99121 battery last?

With three 1.5V AAA batteries, the meter provides approximately 1400 hours of continuous use.

What probe is supplied with the HI99121?

It is supplied with the HI12923 preamplified pH and temperature probe with a conical tip, DIN connector, and 1 m cable.

How do I turn the HI99121 on?

Press the button on the front of the meter; if it does not turn on, check that the batteries are properly installed.