This manual provides instructions on how to use the X'TEN surgical light
This is the user manual for the Maquet X'TEN surgical light, designed for medical illumination during operations. It includes instructions for positioning, cleaning, maintenance, bulb replacement, and troubleshooting. Safety warnings and compliance with CE and ISO standards are emphasized. The manual also covers environmental requirements and maintenance procedures to ensure safe and effective operation.
The X'TEN system is available in single, dual, or triple-arm configurations with optional features like LED ambient lighting and video compatibility. It is intended for use in medical environments and must be operated and maintained by trained personnel. The document provides detailed guidance on handling, sterilization, and servicing to maintain safety and performance.
| Nominal illumination (lx) | 100,000 ± 5% |
|---|---|
| Main light diameter (d50 cm) | 6 (23.6") ± 5% |
| Main light diameter (d50 cm) | 5 (5.9") ± 5% |
| Illumination depth (60%) | 50 (19.7") ± 5% |
| Colour temperature (Ra) | 4,500 ± 5% |
| Colour rendering index | 95 |
| Shadow dilution with one mask | 74% |
| Shadow dilution with two masks | 47% |
| Radiant energy | ≤ 4.4 mW.m-2.lx-1 |
| Mode of operation | Continuous operation |
The ambient temperature must be between -20°C to 60°C, relative humidity 20-75%, and atmospheric pressure 500 to 1060 hPa.
The nominal illumination is 100,000 ± 5% lux.
The color temperature is 4,500 ± 5% K.
The shadow dilution with one mask is 74%.