Maquet SERVO-i Neurally Adjusted Ventilatory Assist (NAVA) Guide

This guide introduces the basic concepts and use of Neurally Adjusted Ventilatory Assist (NAVA).


Maquet SERVO-i Neurally Adjusted Ventilatory Assist (NAVA) Guide - cover page
Brand
Maquet
Document type
User Manual
Language
English
Pages
66
File format
PDF
File size
5.8 MB
Published
18 October, 2023
Updated
18 October, 2023
MD5 checksum
32F8258A61E50AD4FD34BA842015F267

About this manual

This guide explains how to use Neurally Adjusted Ventilatory Assist (NAVA) on the Maquet SERVO-i ventilator, both invasive and non-invasive. It covers the physiological basis of NAVA, including the Edi signal from the diaphragm, setup requirements, and triggering mechanisms. The guide also includes alarm management, weaning examples, and definitions for key terms. It is intended as a companion to the full user and service manuals for the device.

What's inside

  • Introduction
  • Background and basic concepts
  • Invasive ventilation with NAVA
  • Alarms for invasive NAVA
  • The NAVA level – invasive NAVA
  • Example of weaning from invasive NAVA
  • Non invasive ventilation with NAVA
  • Alarms for non invasive NAVA
  • The NAVA level – non invasive NAVA
  • Example of weaning from non invasive NAVA

Frequently asked questions

How is the Edi signal measured in the NAVA modes?

The Edi signal is measured in microvolts at a rate of 62.5 times per second.

When does the ventilator stop offering assist during a breath?

The ventilator stops offering assist when the Edi signal falls to 70% of its peak value.