Maquet HCU 20 and HCU 30 Field Safety Corrective Action

Field Safety Corrective Action for Maquet HCU 20 and HCU 30 heater-cooler units to address bacterial contamination risks and update water quality management…


Maquet HCU 20 and HCU 30 Field Safety Corrective Action - cover page
Brand
Maquet
Document type
Service Manual
Language
English
Pages
4
File format
PDF
File size
45 KB
Published
01 November, 2026
Updated
01 November, 2026
MD5 checksum
1B6C08A3036B0D1A68ADD32AC0F708DE

About this manual

This document outlines a Field Safety Corrective Action (FSCA) issued by Maquet Cardiopulmonary GmbH for Heater-Cooler Units HCU 20 and HCU 30. It addresses potential bacterial contamination in system water and provides updated water quality management procedures to ensure safe operation. The FSCA includes instructions to replace external water hoses, deactivate UV lamps (if applicable), and implement new water management protocols. The document confirms that the devices do not emit contaminated aerosols during operation, based on independent testing.

Users are advised to contact their local Maquet representative for hose replacement, UV lamp deactivation, and training on new procedures. The FSCA also requires acknowledgment and distribution within organizations using the affected devices to ensure compliance and safety.

What's inside

  • Field Safety Corrective Action
  • Water Quality Management
  • HCU 20 and HCU 30 Procedures
  • External Water Hoses Replacement
  • UV Lamp Deactivation
  • Training and Implementation
  • Acknowledgement and Distribution

Frequently asked questions

What is the purpose of the Maquet HCU 20 and HCU 30?

They are used for cooling or warming a patient connected to an extracorporeal perfusion circuit and keeping the required patient temperature constant.

What corrective action is required for the HCU 20 and HCU 30?

Implement new water quality management procedures to maintain clean water circuits and create a 'microbial unfriendly' environment.

What should users do to prevent recontamination?

Replace external water hoses when introducing new procedures and follow a yearly replacement interval.

What is the FSN number for this Field Safety Corrective Action?

The FSN Number is 2017 -10-16.