Field Safety Corrective Action for Maquet HCU 20 and HCU 30 heater-cooler units to address bacterial contamination risks and update water quality management…
This document outlines a Field Safety Corrective Action (FSCA) issued by Maquet Cardiopulmonary GmbH for Heater-Cooler Units HCU 20 and HCU 30. It addresses potential bacterial contamination in system water and provides updated water quality management procedures to ensure safe operation. The FSCA includes instructions to replace external water hoses, deactivate UV lamps (if applicable), and implement new water management protocols. The document confirms that the devices do not emit contaminated aerosols during operation, based on independent testing.
Users are advised to contact their local Maquet representative for hose replacement, UV lamp deactivation, and training on new procedures. The FSCA also requires acknowledgment and distribution within organizations using the affected devices to ensure compliance and safety.
They are used for cooling or warming a patient connected to an extracorporeal perfusion circuit and keeping the required patient temperature constant.
Implement new water quality management procedures to maintain clean water circuits and create a 'microbial unfriendly' environment.
Replace external water hoses when introducing new procedures and follow a yearly replacement interval.
The FSN Number is 2017 -10-16.