This Installation & Operating Manual covers the GESTRA NRGT 26-2 and NRGT 26-2s level transmitters, designed for measuring water levels in pressurized steam and hot-water plants, condensate, and feedwater tanks. The document details the scope of delivery, safety notes, and intended usage, including compliance with EU directives and standards such as EN 61508 for functional safety.
| Measuring range | 0 % to 100 % |
|---|---|
| Output signal | 4-20 mA |
| Safety integrity level | SIL 2 |
| Minimum conductivity | 100 μS/cm |
| Operating temperature limit | 0 bar boiler pressure |
The fluid must have a conductivity of at least 100 μS/cm, as lower values significantly affect the measured capacitance.
No, there is a danger of death due to explosion if the equipment is used in potentially explosive atmospheres.
Only the manufacturer, GESTRA AG, is authorized to repair the equipment.