Stryker Blood/Fluid Warmer Cassettes Recall Notice

Stryker Medical is voluntarily recalling Stryker and Gaymar brand Blood/Fluid Warmer Cassettes.


Stryker Blood/Fluid Warmer Cassettes Recall Notice - cover page
Brand
Stryker
Document type
Service Manual
Language
English
Pages
4
File format
PDF
File size
107 KB
Published
26 October, 2023
Updated
26 October, 2023
MD5 checksum
99CE35A183E6E18D1CF180AC565E1133

About this manual

This document is a recall notice issued by Stryker Medical regarding specific models of Blood/Fluid Warmer Cassettes. The issue involves malformed threads on the female luer fitting, which may occur during the bonding process. The recall affects only certain model numbers and lot codes, and does not impact all Stryker and Gaymar cassettes. Users are instructed to quarantine affected devices, inform users, and complete a customer response form.

The potential hazard is remote, with testing indicating it is highly unlikely for malformed threads to break off. No injuries have been reported. Stryker will arrange collection and replacement of affected products upon receipt of the completed form.

What's inside

  • Recall Notice
  • Affected Models
  • Lot Numbers
  • Product Issue
  • Potential Hazard
  • Actions Needed
  • Customer Response Form

Frequently asked questions

Which models are affected by this recall?

The affected models are D25310NL, D25310CE, D25315CE, and D25340CE.

What is the issue with the cassettes?

Malformed threads on the female luer fitting are caused during the bonding operation between the luer fitting and the tubing.

What actions should users take?

Users should immediately locate and quarantine subject devices, inform users, circulate the notice internally, maintain awareness until actions are completed, complete and return the Customer Response Form, and inform Stryker of any issues or adverse events.

Download PDF (107 KB)