Stryker LIFEPAK 15 Performance Inspection Procedure

This manual details the performance inspection procedures for the Stryker LIFEPAK 15 Monitor/Defibrillator, covering safety checks, ECG testing, and…


Stryker LIFEPAK 15 Performance Inspection Procedure - cover page
Brand
Stryker
Document type
Service Manual
Language
English
Pages
85
File format
PDF
File size
3.2 MB
Published
19 October, 2026
Updated
19 October, 2026
MD5 checksum
CEC67214486F5835D63B78425FE856A1

About this manual

The document serves as a comprehensive guide for performing manual test procedures on the LIFEPAK® 15 Monitor/Defibrillator. It is designed to be used as part of a regularly scheduled preventive maintenance routine or after any repair, replacement, or calibration procedure.

The procedure covers a wide range of diagnostic tests including power management, exterior physical inspection, temperature calibration, CO2 testing, NIBP (Non-Invasive Blood Pressure) verification, and extensive ECG performance checks for 3-, 5-, and 12-lead configurations. It also includes specific instructions for QUIK-COMBO and Standard Paddle defibrillator testing, as well as leakage current tests to ensure electrical safety compliance.

What's inside

  • Scope and Applicability
  • Resource Requirements
  • Test Equipment Verification
  • Power Management Test
  • ECG Performance Testing
  • Defibrillator/Pacing Testing
  • Leakage Current Tests
  • Data Management

Specifications

Energy Range (Analyzer)0 to 450 J
Load Resistance (Analyzer)50 ±1%
Waveforms (Analyzer)Simultaneous 12-lead output
Blood Pressure Accuracy± 1% full scale, ± 1 mmHg
Mains Voltage Range (Analyzer)90 V ac rms to 264 V ac rms

Frequently asked questions

When should the Performance Inspection Procedure be performed?

It should be performed as part of a regularly scheduled preventive maintenance routine, or after any repair, replacement, or calibration procedure.

What is required for all test equipment used in this procedure?

All test equipment must have a current calibration label issued by a certified calibration facility.

What is the acceptable leakage current range for ECG Direct Applied Part tests?

The measured current must be between 5 µA and 45 µA.