This manual details the performance inspection procedures for the Stryker LIFEPAK 15 Monitor/Defibrillator, covering safety checks, ECG testing, and…
The document serves as a comprehensive guide for performing manual test procedures on the LIFEPAK® 15 Monitor/Defibrillator. It is designed to be used as part of a regularly scheduled preventive maintenance routine or after any repair, replacement, or calibration procedure.
The procedure covers a wide range of diagnostic tests including power management, exterior physical inspection, temperature calibration, CO2 testing, NIBP (Non-Invasive Blood Pressure) verification, and extensive ECG performance checks for 3-, 5-, and 12-lead configurations. It also includes specific instructions for QUIK-COMBO and Standard Paddle defibrillator testing, as well as leakage current tests to ensure electrical safety compliance.
| Energy Range (Analyzer) | 0 to 450 J |
|---|---|
| Load Resistance (Analyzer) | 50 ±1% |
| Waveforms (Analyzer) | Simultaneous 12-lead output |
| Blood Pressure Accuracy | ± 1% full scale, ± 1 mmHg |
| Mains Voltage Range (Analyzer) | 90 V ac rms to 264 V ac rms |
It should be performed as part of a regularly scheduled preventive maintenance routine, or after any repair, replacement, or calibration procedure.
All test equipment must have a current calibration label issued by a certified calibration facility.
The measured current must be between 5 µA and 45 µA.