Stryker Trauma & Extremities (T&E) Medical Device Cleaning and Sterilization Instructions

This manual provides instructions for the cleaning, disinfection, inspection, packaging, and sterilization of Stryker Trauma & Extremities (T&E) reusable…


Stryker Trauma & Extremities (T&E) Medical Device Cleaning and Sterilization Instructions - cover page
Brand
Stryker
Category
medical technology
Document type
Service Manual
Language
Chinese
Pages
32
File format
PDF
File size
9.3 MB
Published
15 October, 2026
Updated
15 October, 2026
MD5 checksum
AEE3BD8B347AAD2C0004BA37A2DAA76B

About this manual

This document serves as a general guide for the reprocessing of Stryker Trauma & Extremities (T&E) medical devices, including instruments, implants, and trays. It outlines procedures for handling, cleaning, disinfecting, inspecting, packaging, and sterilizing these devices to ensure they are safe for subsequent use. The manual emphasizes that healthcare facilities are ultimately responsible for executing confirmed procedures and that national regulations regarding cleaning and sterilization take precedence over the manufacturer's suggestions where conflicts arise. This manual is written in Chinese.

What's inside

  • Introduction
  • Handling Instructions
  • Cleaning Methods
  • Disinfection
  • Inspection
  • Packaging
  • Sterilization
  • Storage
  • Manufacturer Information
  • Appendices

Specifications

Cleaning pH RangeNeutral pH, Enzymatic, and Alkaline (pH≤10.9)
Water Quality for RinsingSterile water or water with bacterial count <10/ml
Endotoxin Limit for Rinsing<0.25 EU/ml
Disinfectant Solution TemperatureBelow 40°C

Frequently asked questions

What type of water is recommended for diluting cleaners?

Softened water is recommended for diluting cleaners and disinfectants.

How should soft-shaft instruments be handled during cleaning?

Automated cleaning processes should be prioritized over manual methods for better reproducibility and reliability.

What is the minimum log reduction required for spore testing?

A reduction of at least 3 log cycles is considered acceptable for the cleaning process validation.