This manual provides instructions for the cleaning, disinfection, inspection, packaging, and sterilization of Stryker Trauma & Extremities (T&E) reusable…
This document serves as a general guide for the reprocessing of Stryker Trauma & Extremities (T&E) medical devices, including instruments, implants, and trays. It outlines procedures for handling, cleaning, disinfecting, inspecting, packaging, and sterilizing these devices to ensure they are safe for subsequent use. The manual emphasizes that healthcare facilities are ultimately responsible for executing confirmed procedures and that national regulations regarding cleaning and sterilization take precedence over the manufacturer's suggestions where conflicts arise. This manual is written in Chinese.
| Cleaning pH Range | Neutral pH, Enzymatic, and Alkaline (pH≤10.9) |
|---|---|
| Water Quality for Rinsing | Sterile water or water with bacterial count <10/ml |
| Endotoxin Limit for Rinsing | <0.25 EU/ml |
| Disinfectant Solution Temperature | Below 40°C |
Softened water is recommended for diluting cleaners and disinfectants.
Automated cleaning processes should be prioritized over manual methods for better reproducibility and reliability.
A reduction of at least 3 log cycles is considered acceptable for the cleaning process validation.