This is an operating manual for a micro annular gear pump. It contains information on the pump's application range, product information, technical data, measurements, flow charts, Ex-motor and alternative brushless DC Ex-motor technical data.
This operating manual provides instructions for the integration, operation, and maintenance of HNP Mikrosysteme micro annular gear pumps, specifically the MZR-6359 Ex, MZR-6359X2 Ex, MZR-7259 Ex, and MZR-7259X2 Ex models. The document details the application scope for various liquids, including solvents and oils, and outlines safety requirements for industrial use in Ex-certified environments.
| Displacement volume | 24 µl (MZR-6359) / 48 µl (MZR-7259) |
|---|---|
| Max. system pressure | 80 bar (1160 psi) |
| Flow rate Q | 0.024 - 112 ml/min (MZR-6359) |
| Explosion protection certification | CE II 2G IIC c T4 |
The pumps are suitable for water, watery solutions, solvents, methanol, oils, lubricating liquids, paints, and varnishes.
No, the micro annular gear pumps must not be used for invasive medical applications where liquid is re-introduced to the body.
The minimum dosage volume for the MZR-6359 Ex is 15 µl.
No, they must not be used in aircrafts and spacecrafts or other vehicles without prior consent of the manufacturer.