ee671 is a miniature air flow transmitter produced by e+e company. it operates on the hot-film anemometer principle and features an innovative, very robust e+e sensing element manufactured in thin-film technology combined with innovative transfer-molding. it is dedicated for accurate and reliable measurement in building automation and ventilation applications.


E+E EE671 Miniature Air Flow Transmitter User's Guide - cover page
Brand
E+E
Document type
User Manual
Language
English
Pages
5
File format
PDF
File size
668 KB
Published
21 October, 2023
Updated
21 October, 2023
MD5 checksum
85080B1AEEB987C5B09C7A801E8E8EF1

About this manual

The EE671 is a miniature air flow transmitter based on hot-film anemometer technology, designed for accurate measurement in building automation and ventilation. It features a robust E+E sensing element and supports both analogue and Modbus RTU outputs. Installation requires correct positioning and immersion depth adjustment using a mounting flange or arrow alignment. The device is ESD-sensitive and not surge-proof. Maintenance is minimal under normal conditions, but cleaning with isopropyl alcohol is recommended in polluted environments.

Technical specs include measurement ranges from 0–20 m/s, supply voltage 10–29 V DC, and operating temperature -20 to 60°C. The Modbus RTU protocol supports up to 32 devices with configurable slave ID. Dimensions are 130 x 121 x 80 mm, with IP50/54 protection.

What's inside

  • General
  • Wiring
  • Dimensions
  • Technical Data
  • Mounting
  • Maintenance
  • Modbus RTU
  • Scope of Supply

Specifications

Measurement Range0... 5 m/s (0...1000 ft/min) to 0... 20 m/s (0...4000 ft/min)
Output Signal0 - 1 V (max. 1 mA), 0 - 5 V (max. 1 mA), 0 - 10 V (max. 1 mA), RS485 Modbus RTU
Accuracy±(0.2 m/s / 40 ft/min + 3 % of measured value) at 20 °C (68 °F) / 45 % rh and 1013 hPa (14.7 psi)
Supply Voltage10...29 V DC
Operating Temperature-20...60 °C (-4...140 °F)
Humidity Range5...95 % RH (non-condensing)
Connection TypeCable version: 0.5 m / 2 m PVC cable, 5x0.25 mm²; Plug version: M12 5-pin connector
Protection ClassIP50 (probe head); IP54 (housing)
Response Timeτ90 typ. 4 s
Modbus Slave IDFactory setting: 238 (16-bit integer)

Frequently asked questions

What is the measurement range of the EE671?

The measurement range is 0–5 m/s (0–1000 ft/min) to 0–20 m/s (0–4000 ft/min).

What is the accuracy of the EE671 at 20°C?

Accuracy is ±(0.2 m/s / 40 ft/min + 3 % of measured value) at 20°C (68°F) / 45% RH and 1013 hPa (14.7 psi).

What is the supply voltage requirement?

The supply voltage is 10–29 V DC.

Download PDF (668 KB)

More E+E Manuals


View all E+E manuals →