Philips Respironics CoughAssist E70 User Manual

This document describes how to operate CoughAssist E70, including how to access prescription setting screens, menu access modes and other precautions.


Philips Respironics CoughAssist E70 User Manual - cover page
Brand
PHILIPS Respironics
Category
Other
Document type
User Manual
Language
English
Pages
55
File format
PDF
File size
2.6 MB
Published
20 October, 2023
Updated
20 October, 2023
MD5 checksum
FDB81181B9186B5A8DE158ABF8716E6F

About this manual

The Philips Respironics CoughAssist E70 is a medical device designed to assist patients in loosening, mobilizing, and clearing secretions. It achieves this by providing high-frequency oscillatory vibrations while gradually applying positive pressure to the airway, followed by a rapid shift to negative pressure. The device is intended for use with facemasks, mouthpieces, or adapters for endotracheal and tracheostomy tubes in both hospital and home environments.

What's inside

  • Introduction
  • Therapy Modes and Features
  • Device Setup
  • Viewing and Changing Settings
  • Cleaning and Maintenance
  • Accessories
  • Troubleshooting
  • Technical Specifications

Specifications

Dimensions29.2 cm W x 23.1 cm H x 19.0 cm D
Weight (without battery)3.8 kg (8.4 lbs)
Weight (with battery)4.3 kg (9.4 lbs)
Relative Humidity15 to 95% (non-condensing)
Atmospheric Pressure101 kPa to 77 kPa
Pressure Range-70 to 70 cmH2O
Frequency Range1-20 Hz
Amplitude Range1-10 cmH2O

Frequently asked questions

What are the two levels of menu access on the device?

The device has Full Menu access, which allows changing all settings, and Limited Menu access, which only allows selecting between Presets.

How do you enter Full Menu Access Mode from Limited mode?

From the Standby or Monitor screen, select the Down button and hold the Manual switch to the right (Inhale).

Can the device be used with a facemask or mouthpiece?

Yes, the device may be used either with a facemask or mouthpiece, or with an adapter to a patient's endotracheal or tracheostomy tube.

What is the sound pressure of the device in the Pause phase?

The sound pressure is less than 60 dBA at 1 meter when set at -40 cmH2O/+40 cmH2O.