Instructions include product use, installation and maintenance information
This user manual provides essential guidance for safely installing, wiring, and operating Azbil's CMS0010/0050/0200/0500/1000/2000 gas mass flow meters for hydrogen and helium. It includes safety precautions, mounting instructions, wiring details, operational modes, and troubleshooting steps. The device must be used only with hydrogen or helium gases and requires proper calibration and inert gas purging for hydrogen applications. Users must follow all safety warnings to prevent explosion or damage.
The manual also covers setup procedures including parameter adjustments, zero flow calibration, and totalized flow reset. It specifies compatible fittings and mounting requirements, including horizontal mounting and avoiding vertical installation to prevent measurement drift. Electrical connections must match the receiving device’s input type, and surge protection is recommended for lightning-prone areas.
| Gas Types | Hydrogen and Helium |
|---|---|
| Gas Type Setting | Must be changed for helium or mixed gases |
| Zero Point Calibration | Required before use with hydrogen |
| Maximum Flowrate | Not specified in text |
Check and adjust the flowrate zero point using the actual gas under actual conditions before use.
This device is exclusively for hydrogen and helium gases, and it cannot be used for measurement of gases other than these.
For 1/4 inch connections, the manual recommends using Swagelok Co.,Ltd. SS-400-1-6STSC11 for Swagelok fittings or SS-4-VCR-1-00032SC11 for VCR fittings.
The device must be mounted horizontally because mounting it vertically may cause drift when the flow rate is zero.