Azbil CMS0010/0050/0200/0500/1000/2000 Gas Mass Flow Meter User's Manual

Instructions include product use, installation and maintenance information


Azbil CMS0010/0050/0200/0500/1000/2000 Gas Mass Flow Meter User's Manual - cover page
Brand
Azbil
Document type
User Manual
Language
English
Pages
44
File format
PDF
File size
5.5 MB
Published
20 October, 2023
Updated
20 October, 2023
MD5 checksum
B54A3B5CD2095679ADFF5A936591D9F3

About this manual

This user manual provides essential guidance for safely installing, wiring, and operating Azbil's CMS0010/0050/0200/0500/1000/2000 gas mass flow meters for hydrogen and helium. It includes safety precautions, mounting instructions, wiring details, operational modes, and troubleshooting steps. The device must be used only with hydrogen or helium gases and requires proper calibration and inert gas purging for hydrogen applications. Users must follow all safety warnings to prevent explosion or damage.

The manual also covers setup procedures including parameter adjustments, zero flow calibration, and totalized flow reset. It specifies compatible fittings and mounting requirements, including horizontal mounting and avoiding vertical installation to prevent measurement drift. Electrical connections must match the receiving device’s input type, and surge protection is recommended for lightning-prone areas.

What's inside

  • Safety Precautions
  • Installation
  • Wiring
  • Operations
  • Troubleshooting
  • Specifications

Specifications

Gas TypesHydrogen and Helium
Gas Type SettingMust be changed for helium or mixed gases
Zero Point CalibrationRequired before use with hydrogen
Maximum FlowrateNot specified in text

Frequently asked questions

How should the device be calibrated for hydrogen?

Check and adjust the flowrate zero point using the actual gas under actual conditions before use.

What gases is this meter designed to measure?

This device is exclusively for hydrogen and helium gases, and it cannot be used for measurement of gases other than these.

What type of fittings are recommended for 1/4 inch connections?

For 1/4 inch connections, the manual recommends using Swagelok Co.,Ltd. SS-400-1-6STSC11 for Swagelok fittings or SS-4-VCR-1-00032SC11 for VCR fittings.

How should the device be mounted to prevent measurement errors?

The device must be mounted horizontally because mounting it vertically may cause drift when the flow rate is zero.

Download PDF (5.5 MB)