Hanna Instruments HI5221 & HI5222 pH/mV/ISE/Temperature Bench Meters Instruction Manual

This manual introduces the operation methods and precautions of the HI5221 and HI5222 pH/mV/ISE/Temperature Bench Meters


Hanna Instruments HI5221 & HI5222 pH/mV/ISE/Temperature Bench Meters Instruction Manual - cover page
Brand
Hanna Instruments
Document type
User Manual
Language
English
Pages
84
File format
PDF
File size
3.2 MB
Published
18 October, 2023
Updated
18 October, 2023
MD5 checksum
6A88C68CC66E8129371626053DFE451B

About this manual

This instruction manual covers the operation and setup of Hanna Instruments' HI5221 and HI5222 bench meters, designed for pH, mV, ISE, and temperature measurements. The HI5221 has one channel, while the HI5222 has two channels with independent configurations. Both feature a color LCD, capacitive touch keypad, and USB connectivity for data logging and PC interface. Calibration options include up to five pH buffers and five ISE standards, with automatic or manual temperature compensation. The manual includes setup, measurement, calibration, logging, troubleshooting, and accessory information.

The meters come with standard electrodes and probes, and support additional accessories for extended use. The device is powered by a 12 Vdc adapter and features non-volatile memory to retain settings. The manual also includes warranty and safety information, and contact details for Hanna Instruments’ global support network.

What's inside

  • Preliminary Examination
  • General Description
  • Functional Description
  • Specifications
  • Operational Guide
  • Displaying Modes
  • System Setup
  • pH Setup
  • mV Setup
  • ISE Setup (HI5222 only)

Specifications

Resolution0.1 pH / 0.01 pH / 0.001 pH; 0.1 mV; 0.1 °C
Accuracy @25 °C±0.1 pH / ±0.01 pH / ±0.002 pH; ±1 last significant digit mV; ±0.2 °C
pH CalibrationUp to five points, 8 standard buffers (pH 1.68, 3.00, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45) or 5 custom buffers
DisplayColor Graphic LCD (240 x 320 pixels)
Power Supply12 Vdc adapter
Dimensions160 x 231 x 94 mm (6.3 x 9.1 x 3.7")
Weight1.2 Kg (2.6 lbs.)
Input Channel1 pH/mV (HI5221) or 2 pH/mV/ISE (HI5222)

Frequently asked questions

What accessories come with the meters?

The meters come with HI1131B electrode, HI7662-W temperature probe, HI7082S electrolyte solution, electrode holder, calibration solutions kit, capillary dropper, 12 Vdc power adapter, instruction manual, and quick reference guide.

What are the dimensions and weight of the HI5221 and HI5222 meters?

The meters measure 160 x 231 x 94 mm and weigh 1.2 Kg.

What is the maximum number of data points the meter can store?

The meter can store up to 100,000 total data points.

How long is the warranty for the meters and their electrodes?

The meters are warranted for two years, while electrodes and probes are warranted for six months.