This manual covers the installation and operation instructions for the ETL Water Level Meter.
This manual provides instructions for installing, operating, and maintaining the Geotech ET Water Level Meter, a portable device for measuring water levels in monitoring wells. It includes safety warnings, setup procedures, calibration steps, and maintenance guidelines. The device uses a conductive sensor with adjustable sensitivity to detect water contact, indicated by an audible alert and red LED. Proper storage and battery handling are emphasized to prevent damage or safety hazards.
The manual also covers troubleshooting, specifications, warranty terms, and return policies. It includes diagrams for installation components and compliance with European safety and EMC standards. Users are advised to follow all safety precautions and not use the device with flammable liquids without appropriate interface probes. This manual is written in Spanish.
| Accuracy | 1/100 of a foot or 1 mm |
|---|---|
| Sensor Material | Stainless steel and PTFE |
Lower the probe into the well until it contacts water, then rotate the probe’s dial clockwise until the buzzer and red LED turn off. Adjust sensitivity by rotating the dial clockwise until the buzzer and LED barely activate.
The meter has an accuracy of 1/100 of a foot or 1 millimeter.
No, this device is not designed for use with hydrocarbons or other flammable liquids. Use Geotech’s interface probe line for such applications.
The product is guaranteed for one (1) year from the date of first sale against defects in materials and workmanship.