Plexus C/2500 Service Manual, introduces the maintenance, maintenance and cleaning of the product
This service manual provides detailed instructions for servicing the Plexus C/2500 Aire Select System, including safety precautions, receiving inspection, repair policies, specifications, control panel features, cleaning procedures, routine maintenance, and troubleshooting. It is intended for qualified medical service personnel only. The manual includes test procedures, check sheets, and part lists to guide repairs and ensure proper operation.
The system is designed for medical use and includes safety warnings for flammable anesthetics and electrical hazards. Specifications cover physical dimensions, weight, operating temperature, humidity, electrical requirements, and safety certifications. The control panel features include ON/STANDBY, pressure control, MAX INFLATE, and alternating pressure therapy modes. Cleaning and maintenance involve regular inspection, disinfection, and electrical safety testing.
| Physical Dimensions | 12" x 10" x 5" (30 cm x 25 cm x 13 cm) |
|---|---|
| Weight | 9 lbs. (4.08 kg) |
| Operating Ambient Temperature | 65°F to 75°F |
| Storage Ambient Temperature | 40°F to 105°F |
| Relative Humidity | 30% to 75% Non-condensing |
| Electrical Power | 120 V~ 220 V~ |
| Frequency | 60 Hz 50 Hz |
| Current | 1A 5A |
| IP Protection | IPX0 |
| Agency Approval | TUV IEC 60601.1 (220V only) UL 544, C22.22 No. 125 |
Disconnect power before servicing. Only qualified personnel should repair it. Do not use in the presence of flammable anesthetics. Repairs should be performed only by qualified personnel familiar with repair practices for servicing medical devices.
The Control Unit is warranted for two (2) years and the mattress for one (1) year under the terms of the Plexus Medical warranty in place at the time of purchase.
With the unit in Low Air Loss mode, press the soft/firm keys until one LED bar is lit; verify the manometer reads 9 + 4 mmHg. Adjust to all LED bars lit; verify 27 + 4 mmHg. Press MAX INFLATE; verify 28 + 4 mmHg.