Plexus C/2500 Aire Select System Service Manual

Plexus C/2500 Service Manual, introduces the maintenance, maintenance and cleaning of the product


Plexus C/2500 Aire Select System Service Manual - cover page
Brand
Plexus
Document type
Service Manual
Language
English
Pages
20
File format
PDF
File size
719 KB
Published
18 October, 2023
Updated
18 October, 2023
MD5 checksum
CC81772791832DB9F8B9A793137314A2

About this manual

This service manual provides detailed instructions for servicing the Plexus C/2500 Aire Select System, including safety precautions, receiving inspection, repair policies, specifications, control panel features, cleaning procedures, routine maintenance, and troubleshooting. It is intended for qualified medical service personnel only. The manual includes test procedures, check sheets, and part lists to guide repairs and ensure proper operation.

The system is designed for medical use and includes safety warnings for flammable anesthetics and electrical hazards. Specifications cover physical dimensions, weight, operating temperature, humidity, electrical requirements, and safety certifications. The control panel features include ON/STANDBY, pressure control, MAX INFLATE, and alternating pressure therapy modes. Cleaning and maintenance involve regular inspection, disinfection, and electrical safety testing.

What's inside

  • Safety Precautions
  • Receiving Inspection
  • Repair Policy
  • Specifications
  • Control Panel
  • Cleaning
  • Routine Maintenance
  • Function Testing
  • Troubleshooting
  • Calibration

Specifications

Physical Dimensions12" x 10" x 5" (30 cm x 25 cm x 13 cm)
Weight9 lbs. (4.08 kg)
Operating Ambient Temperature65°F to 75°F
Storage Ambient Temperature40°F to 105°F
Relative Humidity30% to 75% Non-condensing
Electrical Power120 V~ 220 V~
Frequency60 Hz 50 Hz
Current1A 5A
IP ProtectionIPX0
Agency ApprovalTUV IEC 60601.1 (220V only) UL 544, C22.22 No. 125

Frequently asked questions

What are the safety precautions for servicing the Plexus C/2500?

Disconnect power before servicing. Only qualified personnel should repair it. Do not use in the presence of flammable anesthetics. Repairs should be performed only by qualified personnel familiar with repair practices for servicing medical devices.

What is the warranty period for the Plexus C/2500 Control Unit and mattress?

The Control Unit is warranted for two (2) years and the mattress for one (1) year under the terms of the Plexus Medical warranty in place at the time of purchase.

How do I test the pressure settings on the control unit?

With the unit in Low Air Loss mode, press the soft/firm keys until one LED bar is lit; verify the manometer reads 9 + 4 mmHg. Adjust to all LED bars lit; verify 27 + 4 mmHg. Press MAX INFLATE; verify 28 + 4 mmHg.