Instruction manual, introduces the use of Cond 6+ Conductivity/Temp TDS 6+ TDS/Temp Salt 6+ Salinity/Temp, including function introduction, operating instructions, operating methods, etc.
This instruction manual covers the Cond 6+, TDS 6+, and Salt 6+ handheld meters, which are microprocessor-based devices for measuring conductivity, total dissolved solids (TDS), and salinity. The guide serves as both a step-by-step operating guide and a reference for keypad functions, LCD annunciators, and electrode information. It details how to insert batteries, connect the electrode, switch modes, and perform measurements with automatic or manual temperature compensation.
| Conductivity Ranges | 0 to 20.00 µS/cm, 20.0 to 200.0 µS/cm, 200 to 2000 µS/cm, 2.01 to 20.00 mS/cm, 20.1 to 200.0 mS/cm |
|---|---|
| TDS Ranges | 0 to 10.00 ppm, 10.0 to 100.0 ppm, 100 to 1000 ppm, 1.01 to 20.00 ppt, 20.1 up to 200.0 ppt |
| Salinity Resolution | 0.10 to 5.00 % or 1.0 to 50.0 ppt |
| Accuracy | ±1% full scale |
| Temperature Range | -10.0 to 110.0 °C |
| Cell Constant | Selectable 0.1, 1, 10 |
| Temperature Compensation | Automatic / Manual (from 0 to 50 °C) |
The included electrode has a nominal cell constant of k = 1.0.
Yes, the meter allows for manual temperature compensation.
The maximum is 1 calibration point per range.