Eutech Instruments TDS 6+ Manuals

ManualLib has 2 documents for the Eutech Instruments TDS 6+: 2 user manuals. Every file can be read online or downloaded as a PDF for free.

User Manual 2

Key facts

Model
TDS 6+
Documents
2
Languages
EN
Last added
2023-10-18

Start here

Eutech Instruments CON 6+ TDS 6+ Salt 6+ User Manual

Eutech Instruments CON 6+ TDS 6+ Salt 6+ User Manual

User Manual 38 pages · 406 KB · EN

This manual explains how to operate the Eutech Instruments CON 6+, TDS 6+, and Salt 6+ handheld meters for measuring conductivity, TDS, and salinity. It includes setup, calibration, measurement modes, probe care, and troubleshooting. The meters feature automatic temperature compensation, selectable cell constants, and user-adjustable temperature coefficients. Each model has distinct measurement ranges and resolutions, with shared features like a large LCD display and splash-proof keypad.

The manual also covers advanced setup functions, probe maintenance, and how to calculate temperature coefficients and TDS conversion factors. Replacement parts and accessories are listed with ordering codes. Calibration and measurement procedures are detailed with step-by-step instructions and safety notes, including warnings about improper use and battery handling.

Read online

Eutech Instruments TDS 6+ specifications

Conductivity Ranges (Resolution)0 to 20.00 (0.01) µS/cm, 20.0 to 200.0 (0.1) µS/cm, 200 to 2000 (1) µS/cm, 2.01 to 20.00 (0.01) mS/cm, 20.1 to 200.0 (0.1) mS/cm
TDS Ranges (Resolution)0 to 10.00 (0.01) ppm, 10.0 to 100.0 (0.1) ppm, 100 to 1000 (1) ppm, 1.01 to 20.00 (0.01) ppt, 20.1 up to 200.0* (0.1) ppt
Salinity % Resolution0.10 to 5.00 %, 0.01%
Salinity ppt Resolution1.0 to 50.0 ppt, 0.1 ppt
Cell Constant Selectable0.1, 1, 10
Temperature CompensationAutomatic / Manual (from 0 to 50 °C)
Temperature Coefficient0.0 to 3.0% per °C
Normalization Temperature25.0, 20.0 °C (selectable)
TDS factor0.4 to 1.0
Accuracy±1% full scale

Source: Eutech Instruments CON 6+ TDS 6+ Salt 6+ User Manual

All Eutech Instruments TDS 6+ documents by type

User Manual 2

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in these manuals

What are the measurement ranges for the CON 6+ meter?

The CON 6+ measures conductivity in ranges of 0 to 20.00 µS/cm (0.01 resolution), 20.0 to 200.0 µS/cm (0.1 resolution), 200 to 2000 µS/cm (1 resolution), 2.01 to 20.00 mS/cm (0.01 resolution), and 20.1 to 200.0 mS/cm (0.1 resolution).

From: Eutech Instruments CON 6+ TDS 6+ Salt 6+ User Manual

What is the temperature coefficient range?

The temperature coefficient can be set from 0.0 to 3.0% per °C.

From: Eutech Instruments CON 6+ TDS 6+ Salt 6+ User Manual

What is the cell constant of the included electrode?

The included electrode has a nominal cell constant of k = 1.0.

From: Eutech Instruments Cond 6+/TDS 6+/Salt 6+ Handheld Meters User Manual

Can the meter measure temperature without compensation?

Yes, the meter allows for manual temperature compensation.

From: Eutech Instruments Cond 6+/TDS 6+/Salt 6+ Handheld Meters User Manual