Key facts
- Brand
- EUTECH INSTRUMENTS
- Model
- TDS 6+
- Documents
- 2
- Languages
- EN
- Last added
- 2023-10-18
Start here

Eutech Instruments CON 6+ TDS 6+ Salt 6+ User Manual
This manual explains how to operate the Eutech Instruments CON 6+, TDS 6+, and Salt 6+ handheld meters for measuring conductivity, TDS, and salinity. It includes setup, calibration, measurement modes, probe care, and troubleshooting. The meters feature automatic temperature compensation, selectable cell constants, and user-adjustable temperature coefficients. Each model has distinct measurement ranges and resolutions, with shared features like a large LCD display and splash-proof keypad.
The manual also covers advanced setup functions, probe maintenance, and how to calculate temperature coefficients and TDS conversion factors. Replacement parts and accessories are listed with ordering codes. Calibration and measurement procedures are detailed with step-by-step instructions and safety notes, including warnings about improper use and battery handling.
Eutech Instruments TDS 6+ specifications
| Conductivity Ranges (Resolution) | 0 to 20.00 (0.01) µS/cm, 20.0 to 200.0 (0.1) µS/cm, 200 to 2000 (1) µS/cm, 2.01 to 20.00 (0.01) mS/cm, 20.1 to 200.0 (0.1) mS/cm |
|---|---|
| TDS Ranges (Resolution) | 0 to 10.00 (0.01) ppm, 10.0 to 100.0 (0.1) ppm, 100 to 1000 (1) ppm, 1.01 to 20.00 (0.01) ppt, 20.1 up to 200.0* (0.1) ppt |
| Salinity % Resolution | 0.10 to 5.00 %, 0.01% |
| Salinity ppt Resolution | 1.0 to 50.0 ppt, 0.1 ppt |
| Cell Constant Selectable | 0.1, 1, 10 |
| Temperature Compensation | Automatic / Manual (from 0 to 50 °C) |
| Temperature Coefficient | 0.0 to 3.0% per °C |
| Normalization Temperature | 25.0, 20.0 °C (selectable) |
| TDS factor | 0.4 to 1.0 |
| Accuracy | ±1% full scale |
Source: Eutech Instruments CON 6+ TDS 6+ Salt 6+ User Manual
All Eutech Instruments TDS 6+ documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What are the measurement ranges for the CON 6+ meter?
The CON 6+ measures conductivity in ranges of 0 to 20.00 µS/cm (0.01 resolution), 20.0 to 200.0 µS/cm (0.1 resolution), 200 to 2000 µS/cm (1 resolution), 2.01 to 20.00 mS/cm (0.01 resolution), and 20.1 to 200.0 mS/cm (0.1 resolution).
What is the temperature coefficient range?
The temperature coefficient can be set from 0.0 to 3.0% per °C.
What is the cell constant of the included electrode?
The included electrode has a nominal cell constant of k = 1.0.
From: Eutech Instruments Cond 6+/TDS 6+/Salt 6+ Handheld Meters User Manual
Can the meter measure temperature without compensation?
Yes, the meter allows for manual temperature compensation.
From: Eutech Instruments Cond 6+/TDS 6+/Salt 6+ Handheld Meters User Manual